READ[1|2][:SCALar][:POWer:AC]:RATio:RELative? [<expected_value>[,<resolution>[,<source list>]]]
READ[1|2][:SCALar][:POWer:AC]:RATio:RELative?
[<expected_value>[,<resolution>[,<source list>]]]
This command is only relevant for the HP E4419B power meter.
This command sets the specified window's measurement function to ratio
mode with relative mode on, aborts then initiates both channel A and B,
calculates the ratio measurement result using the new sensor data and
places the result on the bus. The result is a ratio based measurement and
is expressed in the units defined by UNIT[1|2]:POWer:RATio. The
relative value used is that set by the
CALCulate:RELative:MAGNitude:AUTO command.
Note
Syntax
READ
1
2
Parameters
Refer to "The Optional Parameters", on page 2-3 for additional details on
the parameters in this command.
HP E4418B/E4419B Programming Guide
INITiate:CONTinuous must be set to OFF on both channels,
otherwise error -213, "INIT ignored" occurs. If TRIGger:SOURce
is set to BUS on either channel, error -214, "Trigger deadlock"
occurs.
:POW
:SCAL
expected_value
Space
DEF
Item
Description/Default
expected_value
The power meter ignores the
(for the
numeric value entered in this
expected
parameter. Any value entered is
power level)
treated like DEF.
:REL
:AC
:RAT
,
resolution
DEF
MEASurement Instructions
?
,
source list
Range of Values
sensor dependent
1
DEF
2-47