READ[1|2][:SCALar][:POWer:AC]? [<expected_value>[,<resolution>[,<source list>]]]
1. The mnemonic
parameter used in the command sub-systems. The parameters must be entered in
the specified order. If parameters are omitted, they will default from the right.
The parameter DEFault is used as a place holder. Specifying
parameter value unchanged.
2. When the measurement result is linear this parameter represents the number
of significant digits. When the measurement result is logarithmic 1 to 4
represents a resolution of 1, 0.1, 0.01 and 0.001 respectively.
Example
Error Messages
INITiate:CONTinuous must be set to OFF, otherwise error -213, "INIT
ignored" occurs.
If TRIGger:SOURce is set to BUS or HOLD, error -214, "Trigger deadlock"
occurs.
If the expected value and resolution parameters are not the same as the
current expected value and resolution settings on the specified window,
error -221, "Settings conflict" occurs.
HP E4418B/E4419B Programming Guide
Item
Description/Default
resolution
A numeric value for the resolution.
If it is unspecified the current
resolution setting is used. If a
value is entered it should
correspond to the current
resolution setting otherwise an
error occurs.
source list
This is the channel on which the
command is implemented.
If unspecified the current window
setup is used. However, on the
HP E4419B, if the window shows a
ratio or difference measurement,
the upper window defaults to
channel A and the lower window to
channel B.
means DEFault. This is not equivalent to the DEFault
DEF
READ2:POW:AC?
MEASurement Instructions
Range of Values
2
1 to 4
1.0, 0.1, 0.01, 0.001
1
DEF
(@1)
(@2) (HP E4419B
only)
DEF
This command queries the lower
window's measurement.
leaves the
2-37