Syntax
SENS
1
SENS2
Parameters
Item
numeric_value
(for CFACtor
and GAIN1)
numeric_value
(for GAIN2)
1. For example, a gain of 60% corresponds to a multiplier of 0.6 and a gain of 150%
corresponds to a multiplier of 1.5.
Example
Reset Condition
On reset, CFACtor|GAIN1 is set to 100% and GAIN2 to 0.0 dB.
HP E4418B/E4419B Programming Guide
[SENSe[1]]|SENSe2:CORRection:CFACtor|GAIN[1|2][:INPut][:MAGNitude]
:
CORR
Description/Default
A numeric value. The default value,
DEF, is 100%. The value for MIN is
1%. The value for MAX is 150%.
A numeric value. The default value,
DEF, is 0.00 dB. The value for MIN is
-100 dB. The value for MAX is
+100 dB.
SENS2:CORR:GAIN2 50
:GAIN
1
2
:CFAC
Space
?
This command sets a channel
offset of 50 dB for channel B.
SENSe Subsystem
:INP
:MAGN
numeric_value
DEF
MIN
MAX
Space
MIN
MAX
Range of Values
1
1 to 150PCT
DEF
MIN
MAX
-100 to +100 dB
DEF
MIN
MAX