Acquire the data

Press the RUN key. The clicking of attenuator relays should be audible. Verify
1 1
resistance readings on the digital multimeter of 50 Ω ± 0.5 Ω (49.5 to 50.5 Ω).
Record the reading in the performance test record.
In the Channel menu change the Coupling field to 1MΩ / DC. The clicking of
2 2
attenuator relays should be audible.
Press the RUN key. Verify resistance readings on the digital multimeter of 1 MΩ ±
3 3
10 kΩ (0.990 to 1.010 MΩ). Record the reading in the performance test record.
In the Channel menu change the Coupling field to 50 Ω/DC and V/Div to 200 mV/Div.
4 4
Repeat steps 1 through 3.
In the Channel menu change the Coupling field to 50 Ω /DC and V/Div to 1 V/Div.
5 5
Repeat steps 1 through 3.
In the Channel menu change the Coupling field to 50 Ω /DC and V/Div to 4 V/Div.
6 6
Repeat steps 1 through 3.
Connect the male end of the BNC tee to the channel 2 input of the oscilloscope
7 7
module.
Repeat from "Set up the logic analyzer" for channel 2, replacing channel 1 with
8 8
channel 2 where applicable.
See Also
If a reading is not within limits, then the attenuator for the out-of-bounds channel should be
replaced ( see chapter 6) .

Perform an operational accuracy calibration

At this point, an operational accuracy calibration should be performed. Follow the
procedure in chapter 4, "To calibrate the oscilloscope."