2. Forms, Records, and Reports

a. Forms, records and reports required for calibration personnel at all levels are
prescribed by TB 750-25.
b. Adjustments to be reported are designated (R) at the end of the sentence in which
they appear. When adjustments are in tables, the (R) follows the designated adjustment.
Report only those adjustments made and designated with (R).
3. Calibration Description. TI parameters and performance applications which pertain
to this calibration are in table 1.
Test instrument parameters
Time base: Option 001 only
Self-check
Signal input: 50Ω input
1 MΩ input
TD-1225A(V)1/U (HEWLETT-PACKARD, MODEL 5342A/H14), TD-1225A(V)2/U (HEWLETT-PACKARD,
MODEL 5342A/H16) AND HEWLETT-PACKARD, MODELS 5342A AND 5343A
Time base:
Option 001 and time base 10544A,
H14 and H16
Self-check
Signal input:
BNC 50Ω position
BNC 1 MHz position
Type N connector TD-1225A(V)1/U and
(5342A) 1 MΩ input connector
APC-3.5 male with collar (SMA compatible)
1
Not checked above 18 GHz.
Table 1. Calibration Description
HEWLETT-PACKARD, MODEL 5340A
Frequency: 10 MHz
Aging rate: <±5 x 10
Line variation: <±1 x 10
Counts and displays 10 MHz for resolution selected
Accuracy: ±1 count, ± time base error
Frequency range: 10 Hz to 18 GHz
Sensitivity: -30 dBm, 10 Hz to 500 MHz
Frequency range: 10 Hz to 250 MHz
Sensitivity: 50 mV rms
Frequency: 10 MHz
Aging rate: <±5 x10
Line variation: <±1 x 10
Counts and displays 75 MHz for resolution selected
Accuracy: ±1 count, ± time base error
Frequency range: 10 Hz to 520 MHz
Sensitivity: 25 mV rms
Frequency range: 10 Hz to 25 MHz
Sensitivity: 50 mV rms
Frequency range: 500 MHz to 18 GHz
Sensitivity: -25 dBm, 500 MHz to 12.4 GHz
Frequency range: 500 MHz to 26.5 GHz
Sensitivity: -33 dBm, 500 MHz to 12.4 GHz
Performance specifications
per day after 24 hour warm-up
-10
for 10% line variation
-10
-35 dBm, 500 MHz to 10 GHz
-25 dBm, 10 to 18 GHz
per day after 24 hour warm-up
-10
for 10% line variation
-10
-20 dBm, 12.4 to 18 GHz
-28 dBm, 12.4 to 18 GHz
-23 dBm, 18 to 26.5 GHz