Position Exchange of the Fine Focus Knob
The fine focus knobs, located on both sides of the main body,
are flat on one end, and convex on the other. The fine focus
knob on each side is attached with a magnet, and can easily be
swapped.
Position them to best suit your usage.
Removing a flat focus knob
A flat fine focus knob can be easily removed by inserting a
hex driver (2 mm across flats) or the likes into the notch in
the knob.
4.3

Resetting the Z-axis Coordinate

The Z-axis coordinate can be reset to zero and used as the
reference point for adjusting the focus. To do so, adjust the
focus on the sample, and then press the Z-RESET button. The
Z-axis coordinate value will be reset to 0.000 um.
Operation on DS-L3 (See "23.2 Controlling the Microscope".)
Chapter 3
Individual Operations
94
Flat fine focus knob
Convex fine focus knob
Position exchange of the focus knob
[Z-RESET] Button used to reset the Z-axis
coordinate