(6-4)
Software Limit
(See DS-L3 Chapter 6 "6.4 Setting the Software Limits".)
Programmatically set the limit position of the microscope's elevating section and the motorized XY stage movement.
In order to set the limit in XY direction, the motorized XY stage must be connected.
Setting items
Z-axis upper limit
X-axis maximum value
X-axis minimum value
Y-axis maximum value
Y-axis minimum value
Drive the unit to where you want it to be the limit and press an
appropriate button to set the limit position.
Restriction of elevating section
 Stop position during actual operation
When operating the elevating section with the focus
knob, the actual stop position is up to 20 μm beyond
the set upper/lower limit.
 Set upper/lower limit position
For focusing stage system, the upper limit is set 0.5μm
above the current position and for focusing nosepiece
system, the lower limit is set 0.5μm below the current
position.
 If the elevating section is currently at the limit
position, it cannot be returned to default (factory
setting).
In this case, move the elevating section below the limit
position (above the limit position for focusing
nosepiece system) and then press the [RESTORE
DEFAULT] button once more.
XY movement restriction
 Stop position during actual operation
When controlling the stage in the XY direction with the
ergo controller or joystick controller, the actual stop
position of the stage is up to 800μm beyond the set
limit.
 Maximum/minimum value to be set
When setting the maximum or minimum value, a value
0.5 μm greater or less than the current position is set
(the front, back, left, and right limits are also the current
value added or deducted by 0.5 μm).
 If the stage is currently at the limit position in the
XY direction, it cannot be returned to default
(factory setting).
In this case, move the stage within the limit position
and then press the [RESTORE DEFAULT] button once
more.
Chapter 3
Individual Operations
Default setting
2000.000 μm
34000.000 μm
-34000.000 μm
27000.000 μm
-27000.000 μm
164
Configurable range
250.000 μm to 2000.000 μm
10000.000 μm to 34000.000 μm
-34000.000 μm to -1000.000 μm
1000.000 μm to 27000.000 μm
-27000.000 μm to -1000.000 μm
Z, X, and Y value settings