Nikon Eclipse Ni-E Instruction Manual page 80

Upright motorized microscope
Hide thumbs Also See for Eclipse Ni-E:
Table of Contents

Advertisement

Differential Interference
Contrast Microscopy
Switch from Differential Interference Contrast Microscopy to Bright-field Microscopy
Loosen the polarizer rotation clamp screw and rotate the
upper part of the rotatable polarizer unit clockwise to
remove the polarizer from the optical path.
(1) If the analyzer slider or lambda plate slider is in optical
path, pull out to the first click-stop position and remove
it from optical path.
Pull out the DIC slider at objective position from
optical path and remove it.
(2)
Operate the motorized universal condenser to bring the
[OPEN] position into the optical path.
Low magnification microscopy
To perform bright-field microscopy with a 2x or 4x
objective, attach the "2-4x auxiliary lens" to the
condenser turret and move it into the optical path.
When doing so, fully open the aperture and field
diaphragms for the maximum range of illumination.
For details on other operations and precautions, see "3
Bright/Dark-field Microscopy Procedure".
Switch from Differential Interference Contrast Microscopy to Epi-Fluorescence Microscopy
See the end of "4 Epi-Fluorescence Microscopy Procedure".
Chapter 2
Microscopy Operations
Preparation
1
2
3
4
5
6
7
8
9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 ■
66
Focus and Optical
Microscopy
System Adjustment
7
5
3
7
D - D P
0
1
JAPAN
Polarizer rotation
clamp screw
Polarizer removed from optical path
Analyzer
slider
Lambda
plate
Objective side DIC
slider
(or dummy slider)
Removing the optical element for differential
interference contrast microscopy
from the optical path
P A
_ _
2 x
_ _ _ _
Z :
_ _
O P E N
_ _
A S 1 2 . 3 m m
10x
1 2 3 5 . 5 5 5 u m
_ _
F S 2 5 . 5 m m