22 Maintenance and Troubleshooting

Maintenance Task and Test Schedule

The following tasks are for Philips-qualified service professionals only. All maintenance tasks and
performance tests are documented in detail in the service documentation supplied on the monitor
documentation CD.
Ensure that these tasks are carried out as indicated by the monitor's maintenance schedule, or as
specified by local laws. Contact a Philips-qualified service provider if your monitor needs a safety or
performance test. Clean and disinfect equipment to decontaminate it before testing or maintaining it.
230
To Purchase, Visit
Avobus.com
Maintenance and Test Schedule
Monitor Tests
Safety checks. Selected tests on the basis
of IEC 60601-1
Monitor Maintenance
Check ECG synchronization of the
monitor and defibrillator (only if
hospital protocol requires use of
monitor during defibrillation)
Measurement Tests
Performance assurance for all
measurements not listed below.
Measurement Maintenance
NBP calibration
Microstream CO
calibration and
2
performance test
Mainstream and sidestream CO
calibration check
Battery Maintenance
Battery
or call
1-800-674-3655
Frequency
At least once every two years, or as needed, after any repairs
where the power supply is removed or replaced, or if the
monitor has been dropped.
At least once every two years, or as needed.
At least once every two years, or if you suspect the measurement
values are incorrect.
At least once every two years, or as specified by local laws.
At least once a year or after 4000 operating hours.
At least once a year, or if you suspect the measurement values are
2
incorrect.
See the section on Maintaining Batteries