Philips IntelliVue X2 Instructions For Use Manual page 271

Multi-measurement module
Hide thumbs Also See for IntelliVue X2:
Table of Contents

Advertisement

Measurement Specifications
NBP
Complies with IEC 60601-2-30:1999/EN60601-2-30:2000.
To Purchase, Visit
Avobus.com
SpO
Alarm
2
Specifications
Range
SpO
Adult: 50 to 100%
2
Pedi/Neo: 30 to 100%
Desat
Adult: 50 to Low alarm limit
Pedi/Neo: 30 to Low alarm limit
Pulse
30 to 300 bpm
Tachycardia
Difference to high limit 0 to 50 bpm 5 bpm steps
Clamping at 150 to 300 bpm
Bradycardia
Difference to low limit 0 to 50 bpm
Clamping at 30 to 100 bpm
NBP Performance Specifications
Measurement Ranges Systolic
Diastolic
Mean
Pulse Rate
Accuracy
Pulse Rate Measurement Accuracy
Pulse Rate Range
Measurement Time
or call
1-800-674-3655
Adjustment
1% steps
1% steps
Adult:
1 bpm steps (30 to 40 bpm)
5 bpm steps (40 to 300 bpm)
Pedi/Neo:
1 bpm steps (30 to 50 bpm)
5 bpm steps (50 to 300 bpm)
5 bpm steps
5 bpm steps
5 bpm steps
Adult: 30 to 270 mmHg (4 to 36 kPa)
Pedi: 30 to 180 mmHg (4 to 24 kPa)
Neo: 30 to 130 mmHg (4 to 17 kPa)
Adult: 10 to 245 mmHg (1.5 to 32 kPa)
Pedi: 10 to 150 mmHg (1.5 to 20 kPa)
Neo: 10 to 100 mmHg (1.5 to 13 kPa)
Adult: 20 to 255 mmHg (2.5 to 34 kPa)
Pedi: 20 to 160 mmHg (2.5 to 21 kPa)
Neo: 20 to 120 mmHg (2.5 to 16 kPa)
Adult: 40 to 300
Pedi: 40 to 300
Neo: 40 to 300
Max. Std. Deviation: 8 mmHg (1.1 kPa)
Max. Mean Error: ±5 mmHg (±0.7 kPa)
±
40 to 100 bpm:
5 bpm
±
101 to 200 bpm:
5% of reading
±10% of reading
201 to 300 bpm:
(average over NBP measurement cycle)
40 to 300 bpm
Typical at HR >60 bpm
Auto/manual/sequence: 30 seconds (adult)
25 seconds (neonatal)
Stat: 20 seconds
Maximum time: 180 seconds (adult/pediatric)
90 seconds (neonates)
24 Specifications
Delay
(0, 1, 2, 3,... 30)
+4 seconds
max. 14 seconds
max. 14 seconds
max. 14 seconds
263

Advertisement

Table of Contents
loading

Table of Contents