Technical Alarm Messages (INOPs)
INOP Message, Indication
C.O. UNPLUGGED
numeric is replaced by -?-
INOP tone.
CO
UPDATE FW
2
Numeric is replaced by a -?-
INOP tone.
CO
WAIT CAL2
2
Numeric is replaced by a -?-
CO
ZERO FAILED
2
Numeric is replaced by a -?-
INOP tone.
CO
ZERO REQU'D
2
Numeric is replaced by a -?-
INOP tone
CO
ZERO RUNNING
2
CPP CHK SOURCES
Numeric is replaced by a -?-
CPP CHK UNITS
Numeric is replaced by a -?-
!!/!!!CUFF NOT DEFLAT
Numeric is displayed with a -?-
Severe yellow/red INOP tone.
During this INOP, alarms cannot be
paused or switched off.
!!/!!!CUFF OVERPRESS
Numeric displayed with -?- ;
Severe yellow/red INOP tone.
During this INOP, alarms cannot be
paused or switched off.
CVP INOPS
DEVICE CHECK SETUP
INOP tone.
DEVICE CHECK CONF.
INOP tone.
DEVICE DEMO DATA
INOP tone
Source
What to do
C.O.
Plug in the C.O. module. Silencing this INOP switches off the
measurement.
CO
The software in the Measurement Extension does not match
2
the software in the MMS. Contact your service personnel.
CO
Calibration on the first calstick cell is complete. Place the
2
transducer on the other calstick cell and start the CAL2
calibration cycle.
CO
An error occurred during the last zero calibration. Check the
2
airway adapter and clean, if necessary. Perform another zero
calibration. If the INOP persists, contact your service
personnel.
CO
Perform zero calibration for the CO
2
persists, contact your service personnel.
CO
Wait until zero calibration is finished.
2
CPP
Not all measurements or values required to perform the
calculation are available. Check the measurement sources.
CPP
The monitor has detected a conflict in the units used for this
calculation. Check the unit settings.
NBP
Remove the cuff from the patient. Make sure that the tubing is
not kinked or twisted and that the correct patient category is
selected. Try repeating the measurement.
You can silence the INOP, but the INOP message remains
visible until the next NBP measurement is started or the
Stop All
SmartKey is selected.
[Adult or pediatric patients: The NBP cuff pressure has exceeded
15mmHg (2kPa) for more than 3 minutes.
Neonatal patients: The NBP cuff pressure has exceeded 5mmHg
(0.7kPa) for more than 90 seconds.]
NBP
The NBP cuff pressure exceeds the overpressure safety limits.
Remove the cuff from the patient. Make sure that the tubing is
not kinked or twisted and that the correct patient category is
selected. Try restarting the measurement.
You can silence this INOP, but the INOP message remains
visible until the next measurement is started or the
All
SmartKey is selected.
PRESS
See <Pressure label> INOPS (under Pressure).
IntelliBridge Device identification completed, but communication could
not be established due to timeout.
IntelliBridge INOP abbreviations may differ slightly depending
on the device category.
IntelliBridge Device identification completed, but communication could
not be established due to error.
IntelliBridge INOP abbreviations may differ slightly depending
on the device category.
IntelliBridge The device connected to the IntelliBridge module reports demo
data but the monitor is not in DEMO mode.