Installation Safety Information

Symbols
Installation Safety Information
WARNING
If multiple instruments are connected to a patient, the sum of the leakage currents may exceed the
limits given in IEC/EN60601-1, IEC60601-1-1, UL60601-1. Consult your service personnel.
Earthing
Combining equipment Combinations of medical equipment with non-medical equipment must comply with
Network Cables
Connectors
The actual placement of boards and configuration of connections for your monitor depends on how
your hardware has been configured. See the symbols table on page 248 to see which symbols are used
to mark the connections.
WARNING
• Connect only medical devices to the ECG output connector socket.
Masimo SET
technology
Silence Alarms
ECG connector
IntelliVue
Instrument
Telemetry wireless
network
electronic equipment (WEEE)
Built-in short
range radio
interface
The monitor must be earthed during operation. The earthing is for functional purposes
and does not provide protection against electric shock. The protection against electric
shock in this device is provided by double and/or reinforced insulation. If a three-wire
receptacle is not available, consult the hospital electrician. Never use a three-wire to two-
wire adapter.
IEC 60601-1-1. Never use a multiple portable socket-outlet or extension cord when
combining equipment unless the socket outlet is supplied specifically for use with that
equipment.
All network cables must be unshielded.
Nellcor OxiMax
compatible
Alarms
Error LED
Always use separate
collection for waste
electrical and
Built-in wireless
network
24 Specifications
LAN connection
indicator for
connection to a wired
network
Alarms Off
Identifies year
2002-
and month of
06
manufacture
Monitor supports 12-
lead ECG
249

Advertisement

Table of Contents
loading

Table of Contents