Safety And Performance Tests - Philips IntelliVue X2 Instructions For Use Manual

Multi-measurement module
Hide thumbs Also See for IntelliVue X2:
Table of Contents

Advertisement

24 Specifications

Safety and Performance Tests

You must observe any national regulations on the qualification of the testing personnel and suitable
measuring and testing facilities. See the maintenance section for a list of required tests. Safety and
performance tests, and what to do if the instrument does not meet these specifications are described in
the Installation and Service guide.
Electromagnetic Compatibility (EMC) Specifications
Take special precautions regarding electromagnetic compatibility (EMC) when using medical electrical
equipment. You must operate your monitoring equipment according to the EMC information
provided in this book. Portable and mobile radio frequency (RF) communications equipment can
affect medical electrical equipment.
Accessories Compliant with EMC Standards
All accessories listed in the accessories section comply, in combination with the monitor, with the
requirements of IEC 60601-1-2:2001 + A1:2004.
WARNING
Using accessories other than those specified may result in increased electromagnetic emission or
decreased electromagnetic immunity of the monitoring equipment.
Electromagnetic Emissions
The monitor is suitable for use in the electromagnetic environment specified in the table below. You
must ensure that it is used in such an environment
Emissions test
Radio Frequency (RF) emissions
RF emissions CISPR 11
for the MP2 or X2 with all accessories
except the M2741A Sidestream CO
sensor
RF emissions CISPR 11
for the MP2 or X2 whenever used with
the M2741A Sidestream CO
Avoiding Electromagnetic Interference (Resp)
The respiration (Resp) measurement is a very sensitive measurement that measures a very small signal.
Technological limitations do not allow higher immunity levels than 1 V/m for radiated RF
electromagnetic fields and 1 Vrms for conducted disturbances induced by RF fields. Electromagnetic
fields with field strengths above 1 V/m and conducted disturbances above 1 Vrms may cause erroneous
measurements. Therefore Philips recommends that you avoid using electrically radiating equipment in
the close proximity of this measurement.
270
Compliance
Avoiding Electromagnetic Interference
Group 1
The monitor uses RF energy only for its internal function.
Therefore, its RF emissions are very low and are not likely
to cause any interference in nearby electronic equipment
Class B
The monitor is suitable for use in all establishments,
including domestic establishments and those directly
connected to the public low-voltage supply network that
2
supplies buildings used for domestic purposes.
Class A
The monitor is suitable for use in all establishments other
than those directly connected to the public low-voltage
power supply network that supplies buildings used for
sensor
2
domestic purposes.
Safety and Performance Tests

Advertisement

Table of Contents
loading

Table of Contents