Measurement Specifications
Respiration Alarm
Specifications
High
Low
Apnea Alarm
Range
Adult/pedi: 10 to 100 rpm
Neo: 30 to 150 rpm
Adult/pedi: 0 to 95 rpm
Neo: 0 to 145 rpm
10 to 40 seconds
Adjustment
Delay
under 20 rpm: 1 rpm steps
max. 14 seconds
over 20 rpm: 5 rpm steps
under 20 rpm: 1 rpm steps
for limits from 0 to 20 rpm:
over 20 rpm: 5 rpm steps
max. 4 seconds
for limits above 20 rpm: max.
14 seconds
5 second steps
24 Specifications