Philips IntelliVue X2 Instructions For Use Manual page 82

Multi-measurement module
Hide thumbs Also See for IntelliVue X2:
Table of Contents

Advertisement

5 Patient Alarms and INOPs
74
To Purchase, Visit
Avobus.com
INOP Message, Indication
PPV BAD SIGNAL
at Information Center
PPV CHK SOURCES
<Pressure label> ARTIFACT
Numeric questionable
<Pressure label>
DEACTIVATED
INOP tone
<Pressure label> EQUIP
MALF
Numeric is replaced by -?-
INOP tone.
<Pressure label> NO
TRANSDUCER
Numeric is replaced by -?-
INOP tone.
<Pressure label> NOISY
SIGNAL
Pulse numeric is replaced by -?-
INOP tone.
<Pressure label> NON-
PULSATILE
Pulse numeric is replaced by -?-
INOP tone.
<Pressure label>
OVERRANGE
Numeric is replaced by -?-
INOP tone.
<Pressure label> REDUCE
SIZE
<Pressure label>
UNPLUGGED
INOP tone.
<Pressure label>
ZERO+CHECK CAL
Numeric is replaced by -?-
or call
1-800-674-3655
Source
What to do
PPV
The arterial pressure source selected for PPV is not providing a
pulsatile signal.
PPV
The arterial pressure source selected for PPV is unplugged or
switched off. When this INOP has displayed for 1 minute PPV
will be switched off.
PRESS
A non-physiological event is detected (for example, a flush or
blood sample). A resulting high limit alarm will be suppressed.
PRESS
A Pressure measurement label in the measurement device or
extension has been deactivated, either by connecting a Temp
transducer in the shared Press/Temp socket, or by deactivating
the label in the Measurement Selection window.
The measurement automatically disappears from the display.
To switch the measurement on again, either reconnect a
Pressure transducer or reactivate the measurement label in the
Measurement Selection window.
PRESS
Contact your service personnel.
The pressure hardware is faulty.
PRESS
Make sure that the pressure transducer is connected to the
measurement device or module.
If you silence this INOP, the measurement will be switched off.
PRESS
This INOP can only arise when a pressure is selected as the
pulse source. It occurs when the pulse detector finds a pulse
rate above 350bpm. This is usually caused by movement
artifact or electrical interference.
PRESS
This INOP can only arise when a pressure is selected as the
pulse source. It occurs when the pulse rate being measured is
less than 25 beats per minute or the amplitude is less than three
mmHg.
Check the catheter and connections to the patient.
PRESS
Make sure that the measurement has been properly prepared
and zeroed, and that the transducer is level with the heart. If
this INOP persists, try another transducer.
Possible causes are a measured pressure outside the allowed
pressure range, or a broken wire to the transducer.
PRESS
Increase the scale for the pressure wave.
PRESS
A Pressure measurement label has been deactivated, either by
unplugging a module, or by deactivating the label in the
Measurement Selection window.
The measurement automatically disappears from the display.
To switch the measurement on again, either replug the module
or reactivate the measurement label in the Measurement
Selection window.
PRESS
Perform a zero and check the calibration of the transducer.
Technical Alarm Messages (INOPs)