Changing Monitor Settings
When the X2 is connected to a host monitor, the Measurement Selection window can be
opened, but only the measurement Setup key is functional. Derived measurements are not active and
cannot be activated, but become active again when the X2 is disconnected from the host monitor.
Resolve any label conflicts at the host monitor.
To resolve a label conflict when the X2 is not connected to a host monitor:
Press the SmartKeys key and
1
– Either select Main Setup and then Meas. Selection
– Or select the Meas. Select. SmartKey
to display the Measurement Selection window.
2
Select the label to be corrected.
3
Use the measurement selection keys to resolve the conflict. Select either:
– Change Label: to assign a different label to the conflicting label.
– De-activate: to disable the conflicting measurement. It retains its label for future use but
becomes invisible to the monitor, as though it had been unplugged.
– Setup <Measurement label>: to enter the Setup menu for the measurement and change
the conflicting device's label to a different label.
Select the De-activate pop-up key to disable the conflicting measurement.
4
Label Compatibility
When a new measurement is introduced, or new labels for an existing measurement, these labels will
not be shown on older Information Centers, and consequently not on the Overview screen sourced
from the Information Center.
When a patient is transferred from a monitor with these new labels to one with an older software
revision, the labels will be replaced with a generic label for that measurement. The settings for that
generic label will then be used.
If it is critical that the measurement labels are available at the Information Center and after transfers,
the older monitors and the Information Center must be upgraded to the appropriate software revision.
Changing Monitor Settings
To change monitor settings such as brightness, or QRS tone volume:
Press the SmartKeys key
1
2
Either Enter the Main Setup menu by selecting the SmartKey
want to change, or select User Interface to enter a submenu where you can change user
interface settings.
Or Select the appropriate SmartKey for the setting you want to change.
Adjusting the Screen Brightness
1
Select the Brightness SmartKey.
.
2 Basic Operation
. Select the setting you
31