Philips IntelliVue X2 Instructions For Use Manual page 78

Multi-measurement module
Hide thumbs Also See for IntelliVue X2:
Table of Contents

Advertisement

5 Patient Alarms and INOPs
INOP Message, Indication
DEVICE REAL DATA
INOP tone
<Device> UNPLUGGED
INOP tone.
DEVICE UNSUPPORTED
INOP tone.
<EC10/EC40> EQUIP MALF
INOP tone.
ECG/ARRH ALARM OFF
!!ECG/AR ALARM OFF
ECG EQUIP MALF
Numeric is displayed with a -?-
INOP tone.
ECG EQUIP MALF T
Numeric is displayed with a -?-
INOP tone.
(!!/!!!)<ECG LEAD> LEAD
OFF
Numeric is displayed with a -?-
INOP tone.
ECG EL. NOISY <ECG LEAD>
(!!/!!!)ECG LEADS OFF
ECG NOISY SIGNAL
INOP tone.
EcgOut EQUIP MALF
INOP tone
EEG EQUIP MALF
INOP tone
EEG IMPEDANCE HIGH or
EEG1 and/or EEG2 IMPED. HIGH
EEG<X> LEAD OFF <n>
[X = channel, n = electrode]
70
Source
What to do
IntelliBridge The monitor is in DEMO mode but the device connected to
the IntelliBridge module reports data that are not flagged as
demo data.
IntelliBridge The IntelliBridge module has been unplugged from the rack, or
the whole rack has been disconnected. Silencing this INOP
switches off the measurement.
IntelliBridge INOP abbreviations may differ slightly depending
on the device category.
IntelliBridge Device identification completed, but no appropriate device
driver installed.
IntelliBridge INOP abbreviations may differ slightly depending
on the device category.
IntelliBridge Malfunction in the IntelliBridge module. If this message
appears repeatedly, the module must be replaced. Contact your
service personnel.
ECG
All ECG alarms have been switched off, or the HR alarm
source is not ECG. To resume ECG alarm generation, switch
ECG alarms on or select ECG as the alarm source.
ECG
Contact your service personnel.
The ECG hardware is faulty.
Monitor
Contact your service personnel.
The ECG in the Telemetry device is faulty.
ECG
Not all the required leads for ECG monitoring are connected.
Check the ECG connections and make sure that the electrode
indicated by <ECG lead> [RA, LA, LL, RL, V or C] electrodes
is attached. In EASI mode, all 5 electrodes must be connected.
ECG
The ECG signal from the named ECG electrodes [RA, LA, LL,
RL, V (or C)] is noisy. Check the ECG connections and make
sure that the electrode indicated is attached.
ECG
Check that all of the required ECG leads are attached, and that
none of the electrodes have been displaced.
ECG
The ECG signal is too noisy. Check that the electrodes are
properly placed and have not dried out. Remove any possible
sources of signal noise (such as power cords) from the area
around the cable and the patient.
The ECG signal may be saturated or overloaded.
ECG
Check that the ECG out cable is securely connected. Contact
your service personnel.
EEG
The EEG hardware is faulty. Contact your service personnel.
EEG
The signal electrode in one or both channels exceeds the user-
selected impedance limit, or the impedance of a single electrode
exceeds the limit. Check the impedance. If the impedance is
too high, reconnect the electrodes according to the EEG
monitoring setup guidelines. If the INOP persists, contact your
service personnel.
EEG
Reconnect specified electrode.
Technical Alarm Messages (INOPs)