Disconnecting from AC Mains Power

To disconnect the X2 from AC mains power, disconnect the MSL cable, or detach the X2 from the
host monitor (if directly connected) or unplug the power cord for the optional external power supply
(M8023A) from the mains socket.

Monitoring After a Power Failure

If external power is disconnected or there is a power failure, the monitor continues to run on its
rechargeable battery.
If the monitor is without any power (no external power or the battery is empty) for less than one
minute, monitoring will resume with all active settings unchanged. If the monitor is without power for
more than one minute, the behavior depends on your configuration. If Automat. Default is set
to Yes, the default profile will be loaded when power is restored. If Automat. Default is set to
No, all active settings are retained, if power is restored within 48 hours. The Automat. Default
setting is made in Configuration Mode.

Networked Monitoring

You can connect your monitor to an Information Center on a network, using one of the optional
interfaces:
• Standard wired LAN
• Wireless LAN
• IntelliVue Instrument Telemetry System (IIT)
WARNING
Do not connect patient monitors to the standard hospital network.
Additionally, when the X2 is equipped with an IIT interface and declared at the Information Center as
a telemetry device, it allows data continuity when paired to a host monitor. After disconnection from
the host monitor, it continues to monitor the patient during transport (or roaming within the specified
area of coverage) and provides continuous data to the same sector on the Information Center. (See the
Instructions for Use for your host monitor for viewing telemetry data on the host.)
If your monitor is connected to a network, a network symbol is displayed in the upper left corner next
to the bed label. To see details about the monitoring equipment and technical information about the
network, select the Main Setup SmartKey to enter the Setup menu, then select
Bed Information.
Be aware that some network-based functions may be limited for monitors on wireless networks in
comparison to those on wired networks.