Technical Alarm Messages (INOPs)

Technical Alarm Messages (INOPs)
If an INOP interrupts monitoring and alarm detection, the measurement numeric will be replaced
by -?-. If an INOP may lead to unreliable measurement values, a ? appears next to the numeric.
INOP Message, Indication
ABP INOPS
ALL ECG ALARMS OFF
Ao INOPS
ART INOPS
AWF CHANGE SCALE
AWP CHANGE SCALE
AWV CHANGE SCALE
Bad Serverlink
INOP tone
BAP INOPS
BATT EMPTY
INOP tone, battery LED flashes
During this INOP, alarms cannot be
paused or switched off.
BATT INCOMPAT
INOP tone
BATT LOW
INOP tone
BATT MALFUNCTION
INOP tone, battery LED flashes
During this INOP, alarms cannot be
paused or switched off unless the
monitor is connected to mains power.
BATTERIES EMPTY or BATT 1/
BATT 2 EMPTY
INOP tone, battery LED flashes
During this INOP, alarms cannot be
paused or switched off.
BATTERIES INCOMP or BATT 1/
BATT 2 INCOMPAT
INOP tone
Source
What to do
PRESS
See <Pressure label> INOPS (under Pressure).
ECG/
All ECG alarms have been switched off, or the HR alarm
Arrhythmia
source is not ECG. To resume ECG alarm generation, switch
ECG alarms on or select ECG as the alarms source.
PRESS
See <Pressure label> INOPS (under Pressure).
PRESS
See <Pressure label> INOPS (under Pressure).
Spirometry
Airway flow signal exceeds range of selected scale. Adjust scale
to display complete wave.
Spirometry
Airway pressure signal exceeds range of selected scale. Adjust
scale to display complete wave
Spirometry
Airway volume signal exceeds range of selected scale. Adjust
scale to display complete wave.
Monitor
1) An MMS with an incompatible software revision is
connected to the monitor. This combination does not allow
monitoring, OR
2) You cannot use this combination of monitor, MMS and
cable. Switch off the monitor and contact your service
personnel.
PRESS
See <Pressure label> INOPS (under Pressure).
Battery
The estimated remaining battery-powered operating time is less
than 10 minutes. Replace the battery immediately.
If the condition persists and the monitor is not connected to
mains power, this INOP is re-issued two minutes after you
acknowledge it.
Battery
The battery cannot be used with this monitor. Replace with the
(M4607A)
correct battery
Battery
The estimated battery-powered operating time remaining is less
than 20 minutes.
Battery
The monitor cannot determine the battery status. If this INOP
persists, replace the faulty battery. If the condition persists and
the monitor is not connected to mains power, this INOP is re-
issued two minutes after you acknowledge it.
Place the battery in a different monitor or in a battery charger.
If the same INOP is shown, contact your service personnel.
Batteries
The estimated remaining battery-powered operating time of
the indicated battery or batteries is less than 10 minutes.
Replace the batteries immediately.
If the condition persists and the monitor is not connected to
mains power, this INOP is re-issued two minutes after you
acknowledge it.
Batteries
The indicated battery or batteries cannot be used with this
monitor. Replace with the correct battery or batteries as
specified in this book.