Monitoring Invasive Pressure

13
WARNING
Measurements from an MMS extension connected to an X2 are not available when the X2 is running
on battery power. They are only available when the X2 is powered from AC mains, either when
connected to a host monitor or the external power supply (M8023A).
CAUTION
Do not use the X2 in combination with monitors using an M1006A pressure module and the
HP1290A pressure transducer. This may cause interference on the respiration or invasive pressure
signals.
You can measure pressure using the monitor's built-in pressure measurement (optional), or one of the
MMS extensions.

Setting up the Pressure Measurement

Plug in the pressure cable.
1
Prepare the flush solution.
2
Flush the system to exhaust all air from the tubing. Ensure that the transducer and stopcocks are
3
free of air bubbles.
WARNING
If air bubbles appear in the tubing system, flush the system with the infusion solution again. Air
bubbles may lead to a wrong pressure reading.
Pressure connector
13
163