Monitoring Carbon Dioxide - Philips IntelliVue X2 Instructions For Use Manual

Multi-measurement module
Hide thumbs Also See for IntelliVue X2:
Table of Contents

Advertisement

Monitoring Carbon Dioxide

14
WARNING
• Measurements from an MMS extension connected to an X2 are not available when the X2 is
running on battery power. They are only available when the X2 is running on external power, either
when connected to a host monitor or the external power supply (M8023A).
• Only one CO
Use the CO
ventilation.
There are two methods for measuring carbon dioxide in the patient's airway:
• Mainstream measurement uses a CO
patient's breathing system. This method is available using the monitor's built-in CO
(optional), the M3014A Capnography Extension or the M3016A Mainstream Extension.
• Sidestream measurement takes a sample of the respiratory gas with a constant sample flow from the
patient's airway and analyzes it with a remote CO
the monitor's built-in CO
or the M3015A Microstream CO
In both cases, the measurement principle is infrared transmission, where the intensity of infrared light
passing the respiratory gas is measured with a photo detector. As some of the infrared light is absorbed
by the CO
measured CO
When using a wet ventilator circuit, monitor mainstream CO
CO
.
2
The partial pressure is calculated from the gas concentration by multiplying the concentration value
with the ambient pressure.
The measurement provides:
• a CO
2
• an end tidal CO
• an inspired minimum CO
• an airway respiration rate (awRR): the number of breaths per minute, calculated from the CO
waveform.
measurement at a time is supported.
2
measurement to monitor the patient's respiratory status and to control patient
2
measurement (optional) or using the M3014A Capnography Extension,
2
molecules, the amount of light passing the gas probe depends on the concentration of the
2
.
2
waveform.
(etCO
) value: the CO
2
2
(imCO
2
sensor attached to an airway adapter directly inserted into the
2
sensor. You can measure sidestream CO
2
Extension.
2
value measured at the end of the expiration phase.
2
): the smallest value measured during inspiration.
2
if available, in preference to sidestream
2
14
measurement
2
using
2
2