Philips IntelliVue X2 Instructions For Use Manual page 187

Multi-measurement module
Hide thumbs Also See for IntelliVue X2:
Table of Contents

Advertisement

Setting up all CO
Suppressing Sampling (not Mainstream CO
To temporarily stop sampling,
Selecting Pump 15Min Off again before fifteen minutes have passed resets the timer to fifteen
minutes.
To re-start the pump,
Changing CO
This refers to CO
You change CO
To switch etCO2 alarms on and off:
To change the etCO
• Touch the high limit on the alarm limits menu button. Choose the high alarm limit from the pop-
• Touch the low limit on the alarm limits menu button. Choose the low alarm limit from the pop-up
To Purchase, Visit
Avobus.com
Measurements
2
Correction
In the Setup CO
O
2
100%, the default value is 20%. If you are not measuring the expired O
(M3014A
5% from the inspired O
only)
subtracting
Humidity
At installation, the monitor is configured to automatically apply either
Pressure Saturated (BTPS) or Ambient Temperature Pressure Dry (ATPD). To see
which, go to the Setup CO
HumidityCorr
N
O
In the Setup CO
2
(M3016A
in the ventilation gas mixture, you must turn this on.
mainstream
If the N
O correction is not available in the Setup CO
2
and M3015A
measurement in your MMS Extension does not require N
microstream
with Gas Corr. (see below).
only)
Gas
In the Setup CO
(M3014A
If Helium or N
only)
appropriate selection.
If the Gas correction is not available in the Setup CO
in your MMS Extension does not require N
correction is setup with N
Agent
In the Setup CO
(M3014A
anesthetic agent (between 0.0% and 20.0%). If an anesthetic agent is present in the
only)
ventilation gas mixture, you must select the appropriate concentration.
in the Setup CO
menu, select Pump 15Min Off or use the CO
2
available.
in the Setup CO
menu, select Pump On.
2
Alarms
2
specific alarms. See the Alarms section for general alarm information.
2
alarm settings in the Setup CO
2
Select etCO
Alarms and toggle between On and Off.
2
alarm limits:
2
up list that opens.
list that opens.
or call
1-800-674-3655
menu, select Oxy. Corr and select a value between 20% and
2
then select the nearest value from the list.
2
menu, and scroll down to look at Hum. Corr
2
.
menu, select N
O Corr. and turn on or off. If N
2
2
menu, select Gas Corr. and select Helium, N
2
O is present in the ventilation gas mixture, you must make the
2
O Corr. (see above).
2
menu, select Agt. Corr. and select the concentration of the
2
2
menu.
2
14 Monitoring Carbon Dioxide
2
Body Temperature
menu, the CO
2
O correction or it is setup
2
menu, the CO
2
2
O or Helium correction or the N
2
)
Pump Off SmartKey if
2
, estimate it
by
or
O is present
2
2
O or turn off.
2
measurement
O
2
179