Philips IntelliVue X2 Instructions For Use Manual page 270

Multi-measurement module
Hide thumbs Also See for IntelliVue X2:
Table of Contents

Advertisement

24 Specifications
SpO
2
Complies with EN ISO 9919:2005 (except alarm system; alarm system complies with IEC 60601-2-
49:2001).
Measurement Validation: The SpO
blood sample reference measured with a CO-oximeter. Pulse oximeter measurements are statistically
distributed, only about two-thirds of the measurements can be expected to fall within the specified
accuracy compared to CO-oximeter measurements. Display Update Period: Typical: 2 seconds,
Maximum: 30 seconds. Maximum with NBP INOP suppression on: 60 seconds.
SpO
Performance Specifications
2
SpO
2
The specified accuracy
is the root-mean-
square (RMS)
difference between the
measured values and
the reference values
Pulse
Sensors
Pulse Oximeter Calibration Range
262
accuracy has been validated in human studies against arterial
2
Range
0 to 100%
Accuracy
Philips Reusable Sensors:
M1191A, M1191AL, M1191ANL, M1191B, M1191BL, M1192A,
M1192AN = 2% (70% to 100%)
M1193A, M1193AN, M1194A, M1194AN, M1195A, M1195AN,
M1196A = 3% (70% to 100%)
M1191T, M1192T, M1193T (Adult), M1196T = 3% (70% to 100%)
M1193T (Neonate) = 4% (70% to 100%)
Philips Disposable Sensors with M1943A(L):
M1132A, M1133A (adult/infant) = 2%
M1901B, M1902B, M1903B, M1904B, M1131A, M1133A (neonate)
= 3% (70% to 100%)
NellcorPB
MAX-A, MAX-AL, MAX-P, MAX-I, MAX-N, D-25, D-20, I-20, N-
25, OxiCliq A, P, I, N = 3% (70% to 100%)
Masimo Reusable Sensors
LNOP DC-I, LNOP DC-IP, LNOP YI, LNCS DC-1, LNCS DC-IP:
2% (70% to 100%)
LNOP TC-I, LNCS TC-I: 3.5% (70% to 100%)
Masimo Disposable Sensors
LNOP Adt, LNOP Adtx, LNOP Pdt, LNOP Pdtx, LNOP Inf-L,
LNCS Adtx, LNCS Pdtx, LNCS Inf-L: 2% (70% to 100%)
LNOP Neo-L, LNOP NeoPt-L, LNCS Neo-L, LNCS NeoPt-L: 3%
(70% to 100%)
Resolution
1%
Range
30 to 300 bpm
Accuracy
±2% or 1 bpm, whichever is greater
Resolution
1 bpm
Wavelength range: 500 to 1000 nm
Emitted Light Energy: ≤15 mW
Information about the wavelength range can be especially useful to
clinicians (for instance, when photodynamic therapy is performed)
70% to 100%
Measurement Specifications
®
Sensors with M1943A(L):
®
with LNOP MP12 or LNC MP10:
®
with LNOP MP12 or LNC MP10: