Battery Power Indicators
Battery status/malfunction indicator: Normal battery function is indicated by the battery power
gauge, together with the remaining operating time, on the Main Screen. You are informed of problems
or changes in the status of the battery by the battery status/malfunction indicator. This consists of a
blank battery gauge containing a symbol. If the symbol is red, this indicates a critical situation. You can
check the specific cause of the problem by looking at the symbol(s) displayed in the Battery
Status window (see page 223).
Battery status indicator
Monitoring Time Available: While the X2 is running on battery power, a time is displayed below the
battery power gauge. No time is displayed when the X2 is running on external power via a host
monitor or the external power supply (M8023A). This is the estimated monitoring time available with
the current battery power. Note that this time fluctuates depending on the system load (the display
brightness and how many measurements you carry out), the age of the battery, and the remaining
capacity of the battery.
Battery Status Window
To access the Battery Status window and its associated pop-up keys, select the battery status
information on the Screen, or select Main Setup -> Battery.
Capacity, remaining tells you how much power is left in the battery.
Capacity, fullCharge tells you how much power the battery can hold when fully charged.
Alternates with the battery gauge on the
Main Screen.
!
Battery Status
Check in the
window to see which status symbol is
displayed to identify the cause.
Battery Status
TimeToEmpty:
Capacity
remaining
fullCharge
Battery malfunction indicator
The red
!
situation or malfunction. Check
in the
window to see which malfunction
indicator is displayed, or refer to
the INOP, to identify the cause.
Indicator for missing battery
X
(flashing red X). An INOP is
issued when the battery
compartment is empty, and the
X2 is connected to external power
(a host monitor or the optional
external power supply). This
!!INSERT BATTERY
INOP is suppressed for 15
seconds while the monitor is
connected to AC mains power,
allowing you sufficient time to
load a new battery. After
silencing, the INOP reappears
every 10 seconds until a battery is
loaded.
2:33 hrs
[mAh]
[mAh]
21 Using Batteries
!
flashes. Critical battery
Battery Status
223

Advertisement

Table of Contents
loading

Table of Contents