Philips IntelliVue X2 Instructions For Use Manual page 89

Multi-measurement module
Hide thumbs Also See for IntelliVue X2:
Table of Contents

Advertisement

Technical Alarm Messages (INOPs)
INOP Message, Indication
!!/!!!TELE DISCONNECT
INOP tone
TELE EQUIP MALF
INOP tone
TELE INCOMPATIBLE
!!/!!! TELE INOP
Severe yellow/red INOP tone.
Tele Sync Unsupp.
INOP tone
TELE UNSUPPORTED
INOP tone
Δ
<
Temp> CHK SOURCES
Numeric is replaced by -?-
Δ
<
Temp> CHK UNITS
Numeric is replaced by -?-
<Temp label> DEACTIVATED
INOP tone
<Temp label> EQUIP MALF
Numeric is replaced by -?-
INOP tone.
<Temp label> NO
TRANSDUCER
Numeric is replaced by -?-
INOP tone.
<Temp label> UNPLUGGED
INOP tone
Source
What to do
Telemetry
Telemetry transceiver was disconnected or short range radio
link was lost.
For cable connections; check Telemetry interface, cable
connection and setup.
For short range radio connections: if the telemetry transceiver
has not moved out-of-range, check for interference sources
close to the monitor (bluetooth devices, DECT phones,
cellular phones, microwaves, etc.) If this INOP persists, ask
your service personnel to survey the interference sources.
Monitor
The telemetry device has a malfunction. Disconnect and
reconnect the telemetry device. If the INOP reappears, replace
the telemetry device.
Monitor
SRR-enabled telemetry device is not supported by this central
software revision. Please check configuration.
Telemetry
Check for further details at the Information Center or in the
Telemetry Data window on the monitor.
Telemetry
The MMS in use does not support synchronization of ECG
and SpO
settings between the monitor and central station
2
after a telemetry device has been paired. Use an MMS with
revision E.0 or above.
Monitor
This telemetry device is not supported for direct connection to
the monitor.
TEMP
Not all measurements or values required to perform the
Difference
calculation are available. Check measurement sources.
TEMP
The monitor has detected a conflict in the units used for this
Difference
calculation. Check the unit settings.
TEMP
A Temp measurement label in the measurement device has
been deactivated, either by connecting a Pressure transducer in
the shared Press/Temp socket, or by deactivating the label in
the Measurement Selection window.
The measurement automatically disappears from the display.
To switch the measurement on again, either reconnect a Temp
transducer or reactivate the measurement label in the
Measurement Selection window.
TEMP
Contact your service personnel.
The temperature hardware is faulty.
TEMP
Make sure the TEMP probe is connected to the MMS or
module.
If you silence this INOP, the measurement will be switched off.
TEMP
A Temp measurement label has been deactivated, either by
unplugging a module, or by deactivating the label in the
Measurement Selection window.
The measurement automatically disappears from the display.
To switch the measurement on again, either replug the module
or reactivate the measurement label in the Measurement
Selection window.
5 Patient Alarms and INOPs
81