Measurement Specifications
Invasive Pressure and Pulse
Complies with IEC 60601-2-34:2000/EN60601-2-34:2000.
Invasive Pressure Performance Specifications
Measurement Range
Pulse Rate
Input Sensitivity
Transducer
Frequency Response
Zero Adjustment
Gain Accuracy
Overall Accuracy
Volume displacement of CPJ840J6
Invasive Pressure Alarm Specifications Range
Pressure
Extreme High
Extreme Low
Pulse
–40 to 360 mmHg
Range
25 to 350 bpm
Accuracy
±1% Full Range
Resolution
1 bpm
Sensitivity:5 µV/V/mmHg (37.5 µV/V/kPa)
Adjustment range:±10%
Load Impedance:200 to 2000 Ω (resistive)
Output Impedance:≤3000 Ω (resistive)
dc to 12.5 Hz or 40 Hz
Range:
±200 mmHg (±26 kPa)
Accuracy
±1 mmHg (±0.1 kPa)
Drift
Less than 0.1 mmHg/°C (0.013 kPa/°C)
Accuracy
±1%
Drift
Less than 0.05%/°C
Error of ≤0.4% FS (@CAL 200 mmHg)
Non linearity
and Hysteresis
(including
±4% of reading or ±4 mmHg (±0.5 kPa), whichever is greater
transducer)
0.1 mm
–40 to 360 mmHg
(–5.0 to 48 kPa)
Difference to high limit 0 to
25 mmHg
Clamping at -40 to
360 mmHg
Difference to low limit 0 to
25 mmHg
Clamping at -40 to
360 mmHg
25 to 300 bpm
3
/100 mmHg
Adjustment
-40 to 50 mmHg
2 mmHg (0.5 kPa)
>50 mmHg
5 mmHg (1 kPa)
5 mmHg steps (0.5 kPa)
5 mmHg steps (1.0 kPa)
5 mmHg steps (0.5 kPa)
5 mmHg steps (1.0 kPa)
Adult:
1 bpm steps (25 to 40 bpm)
5 bpm steps (40 to 300 bpm)
Pedi/Neo:
1 bpm steps (25 to 50 bpm)
5 bpm steps (50 to 300 bpm)
24 Specifications
Delay
max.
12 seconds