5 Patient Alarms and INOPs
INOP Message, Indication
LA LEAD OFF
Numeric is replaced by -?- for 10
seconds; INOP tone.
LAP INOPs
LEADSET UNPLUGGED
LL LEAD OFF
Numeric is replaced by -?- for 10
seconds; INOP tone.
MCC Reversed
INOP tone
MCC Unsupported
INOP tone
Meas. DEACTIVATED
MEASSRV UNSUPPORTD
INOP tone
MMS Ext. EQUIP MALF
INOP tone
MMS Ext. UNPLUGGED
INOP tone
MMS Ext. Unpowered
INOP tone
MMSExt.Unsupported
INOP tone
MMS UNPLUGGED
INOP tone.
MMS UNSUPPORTED
INOP tone.
!!/!!!MORE BED ALARMS
At Information Center
MSL Power High
MSL Power Off
INOP tone.
MSL Power Overload
INOP tone.
72
Source
What to do
ECG
The LA electrode has become detached from the patient or the
lead set has been changed. Reattach the electrode or select
Lead Setup
new lead set.
PRESS
See <Pressure label> INOPS (under Pressure).
Telemetry
The leadset has been unplugged from the telemetry device.
ECG
The LL electrode has become detached from the patient or the
lead set has been changed. Reattach the electrode or select
Lead Setup
new lead set.
Monitor
The MSL coupling cable is reversed. Connect the end with the
grey connector to the Intelligent Display.
Monitor
An MSL coupling cable has been connected to a device which
does not support MSL coupling.
An X2 or MP5 has been connected to a host monitor
(companion mode) and all derived measurements have been
deactivated and/or measurements with a label conflict. The
measurements can only be reactivated by disconnecting the
measurement device from the host monitor.
MMS
The Multi-Measurement module is not supported by the
monitor. Contact your service personnel.
MMS
Loss of communication between the Multi-Measurement
Extension
Module and the MMS extension. Contact your service
personnel.
MMS
The MMS extension has been disconnected from the Multi-
Extension
Measurement Module.
MMS
The MMS extension cannot operate while the Multi-
Extension
Measurement Module is running on battery power.
MMS
The MMS extensions not supported by your monitor. Contact
Extension
your service personnel.
MMS
Make sure that the Multi-Measurement Module is connected
to the monitor. All MMS measurements are off while the MMS
is unplugged.
MMS
The Multi-measurement Module is not supported by your
monitor. Contact your service personnel.
Monitor
The monitor is associated with a telemetry device and is
sending data to the Information Center via the telemetry
device. There are currently more alarms at the bedside than can
be transmitted to the Information Center.
Monitor
The power consumption of the devices connected to the
Measurement Link (MSL) cable is too high. If this situation
continues, the MSL will be switched off. Contact your service
personnel.
Monitor
The power consumption of the devices connected to the
Measurement Link (MSL) cable was too high for too long and
the MSL has been switched off. Contact your service personnel.
Monitor
The power consumption of the devices connected to the
Measurement Link (MSL) cable is much too high or there has
been a short circuit. The MSL has been switched off. Contact
your service personnel.
Technical Alarm Messages (INOPs)