Technical Alarm Messages (INOPs)
INOP Message, Indication
NBP Deactivated
INOP tone.
NBP EQUIP MALF
Numeric is replaced by -?-
INOP tone.
NBP INTERRUPTED
Numeric is replaced by -?-
INOP tone.
NBP MEASURE FAILED
Numeric may be displayed with a -?-
INOP tone.
No Central Monit.
INOP tone
!!/!!! NO DEVICE DATA
NO DEVICE DATA
NO ECG AT CENTRAL
NO ECG SOURCE
NO PPV FROM MeasSrv
at Information Center
NO PPV FROM <Device>
OUT OF AREA
P/P1/P2/P3/P4 INOPS
PAP INOPS
PPV BAD <Pressure Label>
SIGNAL
Source
What to do
NBP
The NBP measurement label in the measurement device has
been deactivated by deactivating the label in the Measurement
Selection window. The measurement automatically disappears
from the display. To switch the measurement on again,
reactivate the measurement label in the Measurement Selection
window.
NBP
Remove the cuff from the patient. The NBP hardware is faulty.
Contact your service personnel.
You can silence this INOP, but the INOP message remains
visible until the next measurement is started or the
All
SmartKey is selected.
NBP
Check the tubing and cuff for leakages or kinks. Check that
you are using the correct cuff size and placement, and that the
correct patient category is selected. Try restarting the
measurement.
If the INOP occurs repeatedly, contact your service personnel.
You can silence this INOP, but the INOP message remains
visible until the next measurement is started or the
All
SmartKey is selected.
This INOP arises when the measurement needed longer than
the maximum time for inflation, deflation or the total
measurement.
NBP
Check that you are using the correct cuff size and placement,
and that the correct patient category is selected. Try restarting
the measurement.
You can silence this INOP, but the INOP message remains
visible until the next measurement is started or the
All
SmartKey is selected.
Check the condition and suitability of the patient for NBP
monitoring. Use another cuff to continue measuring.
Monitor
There is a problem with the communication to the network.
Central monitoring is currently not possible (no patient alarms
or information). Check the connection. Contact your service
personnel.
IntelliBridge Communication with connected device has been lost.
Monitor
The ECG measured with the monitor ECG is not being sent to
the Information Center via the telemetry device.
Telemetry
A telemetry device is paired with the monitor but the
Information Center is not detecting a valid ECG signal from
either of them.
MMS or
The measurement device does not supply a beat-to-beat arterial
FMS
pressure value. Contact your service personnel.
MMS or
The measurement device does not supply a beat-to-beat arterial
FMS
pressure value. Contact your service personnel.
Telemetry
The telemetry device has left the access point coverage area.
PRESS
See <Pressure label> INOPS (under Pressure).
PRESS
See <Pressure label> INOPS (under Pressure).
PPV
The arterial pressure source selected for PPV is not providing a
pulsatile signal.
5 Patient Alarms and INOPs