Measurement Specifications
Interfering Gas and Vapor Effects On CO
The specified deviations for M3002A/M3014A are valid when the appropriate corrections are switched
on and set correctly.
To Purchase, Visit
Avobus.com
CO
Alarm
2
Specifications
Range
etCO
High
20 to 95 mmHg (2 to 13 kPa) 1 mmHg (0.1 kPa)
2
etCO
Low
10 to 90 mHg (1 to 12 kPa)
2
imCO
High
2 to 20 mmHg
2
(0.3 to 3.0 kPa)
awRR High
Adult/pedi: 10 to 100 rpm
Neo: 30 to 150 rpm
awRR Low
Adult/pedi: 0 to 95 rpm
Neo: 0 to 145 rpm
Apnea delay
10 to 40 seconds
Gas Level
Gas or Vapor
(% volume fraction)
Nitrous Oxide
60
Halothane
4
Enflurane
5
Isoflurane
5
Sevoflurane
5
Xenon
80
Helium
50
Metered dose inhaler
-
propellants
Desflurane
15
Ethanol
0.1
Isopropanol
0.1
Acetone
0.1
Methane
1.0
or call
1-800-674-3655
Adjustment
Delay
M8102A/M3014A/M3016A: less than
14 seconds
M3015A: less than18 seconds.
steps of 1 mmHg
M8102A/M3014A/M3016A: less than
(0.1 kPa)
14 seconds
M3015A: less than18 seconds.
under 20 rpm: 1 rpm
M8102A/M3014A/M3016A: less than
steps
14 seconds
over 20 rpm:5 rpm
M3015A: less than18 seconds.
steps
M3015A:
settings <20 rpm: less than 8 seconds
>20 rpm: less than 18 seconds
M8102A/M3014A/M3016A
settings <20 rpm: less than 4 seconds
>20 rpm: less than 14 seconds
5 second steps
set apnea delay time +4 seconds
(M8102A/M3014A/M3016A) or
8 seconds (M3015A)
Measurement Values
2
M3002A/M3014A
Additional deviation due to gas interference, measured
at 0 - 40 mmHg CO
±1 mmHg
±2 mmHg
±2 mmHg
±2 mmHg
±2 mmHg
-5 mmHg
±1 mmHg
not specified for use
+5 mmHg
±1 mmHg
±1 mmHg
±1 mmHg
±1 mmHg
24 Specifications
M3015A
2
±5 %
±5 %
±5 %
±5 %
±5 %
±5 %
±5 %
±5 %
±5 %
±5 %
±5 %
±5 %