5 Patient Alarms and INOPs
INOP Message, Indication
SPIRO NO BREATH
SPIRO GAS COMPENS?
SPIRO PATIENT CAT.
Δ
SpO
CHK SOURCES
2
Numeric is replaced by -?-
Δ
SpO
CHK UNITS
2
Numeric is replaced by -?-
<SpO
label> DEACTIVATED
2
INOP tone
<SpO
label> EQUIP MALF
2
Numeric is replaced by -?-
INOP tone.
<SpO
label> ERRATIC
2
Numeric is replaced by -?-
INOP tone.
<SpO
label> EXTD. UPDATE
2
Label is displayed with a -?-
(questionable numeric)
<SpO
label> INTERFERNCE
2
Numeric is replaced by -?-
INOP tone.
<SpO
label> LOW PERF
2
Label is displayed with a -?-
(questionable numeric)
<SpO
label> NOISY SIGN.
2
Numeric is replaced by -?-
INOP tone.
<SpO
label> NON-PULSAT.
2
Numeric is replaced by -?-
INOP tone.
<SpO
label> NO SENSOR
2
Numeric is replaced by -?-
INOP tone.
<SpO
LABEL> POOR SIGNAL
2
Label is displayed with a -?-
(questionable numeric)
78
Source
What to do
Spirometry
No breath was detected for more than 25 seconds. Breath
derived numerics are not available.
Spirometry
Gas compensation is set to "gas analyzer" but not all gases
necessary for compensation are measured by a gas monitor.
Some of the fall-back values provided by the user are used.
Measurement accuracy might be reduced.
Spirometry
Mismatch of patient size configured in the host monitor and
sensor type plugged into the module. Check the instructions
on selecting the correct sensor in the Spirometry chapter.
SpO
Not all measurements or values required to perform the
2
Difference
calculation are available. Check measurement sources.
SpO
The monitor has detected a conflict in the units used for this
2
Difference
calculation. Check the unit settings.
SpO
The SpO
measurement label in the measurement device has
2
2
been deactivated by deactivating the label in the Measurement
Selection window. The measurement automatically disappears
from the display. To switch the measurement on again,
reactivate the measurement label in the Measurement Selection
window.
SpO
The MMS or module is faulty. Unplug and replug the MMS or
2
module. If the INOP persists, contact your service personnel.
SpO
Check the sensor placement. Try another adapter cable and
2
sensor. If the INOP persists, contact your service personnel.
SpO
The update period of displayed values is extended due to an
2
NBP measurement on the same limb or an excessively noisy
signal.
SpO
There is too much interference, caused by a high level of
2
ambient light and/or electrical interference. Cover the sensor to
minimize ambient light. If the INOP persists, make sure that
the sensor cable is not damaged or positioned too close to
power cables.
SpO
Accuracy may be compromised due to very low perfusion.
2
Stimulate circulation at sensor site. If INOP persists, change
the measurement site.
SpO
Excessive patient movement or electrical interference is causing
2
irregular pulse patterns. Try to reduce patient movement or to
relieve the cable strain on the sensor.
SpO
Check the perfusion at measurement site. If necessary,
2
stimulate circulation or change measurement site. If the INOP
is due to NBP measurement on the same limb, wait until the
NBP measurement is finished.
SpO
Make sure the SpO
2
try another adapter cable and sensor. If you silence this INOP,
the measurement will be switched off.
SpO
The signal condition of the SpO
2
measurement accuracy may be compromised.
Technical Alarm Messages (INOPs)
sensor is connected. If the INOP persists,
2
measurement is poor and
2