2 Basic Operation
Any labels already being used in the monitor are shown "grayed-out" in the list and cannot be selected.
Resolving Label Conflicts
Each label must be unique, that is, it can only be assigned once. If you have a MMS Extension
equipped with a pressure measurement connected to the monitor, there is a potential conflict with, for
example, the ABP label. If you manually enter measurement values these may also conflict with existing
labels on the monitor.
Depending on your configuration, the monitor will either
• display the Measurement Selection window automatically
for you to resolve the conflict
• take no action, you must enter the Measurement Selection
window and resolve the conflict
All the currently available measurements are depicted in the
Measurement Selection window. Any measurement labels
causing a label conflict are shown in red. If a measurement is connected but currently unavailable, for
example, because it was deactivated due to a label conflict, that measurement is shown "grayed-out". If
a MMS Extension is not available, for example if monitor is running on battery power and not an
external power source, the MMS Extension is not displayed.
PAP ZERO+CHECK CAL
Change
Label
On an X2 display, a MMS Extension is only shown in the Measurement Selection window
when the X2 is connected to a host monitor or the external power supply (M8023A), and not when
running on battery power.
WARNING
When an X2 with an active measurement, say SpO
measurement already active, the SpO
DEACTIVATED INOP is displayed. The measurement can only be reactivated if the X2 is
disconnected from the host monitor. The label conflict can be resolved on the host monitor like any
other label conflict.
30
*** APNEA
Measurement Selection
ABP
Tcore
Temp
NBP
PAP
CO
SpO
2
2
ECG
Resp
Setup SpO
De-
Activate
2
CPP
Temp
More
2
, is connected to a host monitor with the same
2
measurement on the X2 is deactivated and the SpO
Using Labels
measurement selection key
Unavailable measurements
are grayed-out
2