SpO
Signal Quality Indicator (Fast SpO
2
WARNING
Inspect the application site every two to three hours to ensure skin quality and correct optical
alignment. If the skin quality changes, move the sensor to another site. Change the application site at
least every four hours.
CAUTION
• Injected dyes such as methylene blue, or intravascular dyshemoglobins such as methemoglobin and
• Interference can be caused by:
SpO
Signal Quality Indicator (Fast SpO
2
The SpO
is available) which gives an indication of the reliability of the current values.
The level to which the triangle is filled shows the quality of the signal; the indicator below shows a
medium signal quality, the signal quality is at a maximum when the triangle is completely filled.
Assessing a Suspicious SpO
Traditionally, pulse rate from SpO
the SpO
because the correct calculation of SpO
When pulse rate is very low, or strong arrhythmia is present, the SpO
the heart rate calculated from ECG but this does not indicate an inaccurate SpO
If you doubt the measured SpO
perfusion indicator instead to assess the signal quality.
N O T E
With pulse oximetry, sensor movement, ambient light (especially strobe lights or flashing lights) or
electromagnetic interference can give unexpected intermittent readings when the sensor is not attached
to a patient. Especially bandage-type sensor designs are sensitive to minimal sensor movement that
might occur when the sensor is dangling.
To Purchase, Visit
Avobus.com
only)
2
carboxyhemoglobin may lead to inaccurate measurements.
– High levels of ambient light or strobe lights or flashing lights (such as fire alarm lamps). (Hint:
cover application site with opaque material.)
– Electromagnetic interference.
– Excessive patient movement and vibration.
numeric is displayed together with a signal quality indicator (if configured and enough space
2
SpO
2
reading. With newer algorithms, such as FAST-SpO
2
or call
1-800-674-3655
Reading
2
was compared with heart rate from ECG to confirm the validity of
2
is not directly linked to the correct detection of each pulse.
2
, use the signal quality indicator (if available) or the pleth wave and
2
10 Monitoring SpO
only)
2
, this is no longer a valid criteria
2
/Pleth pulse rate may differ from
2
value.
2
2
149