Using Batteries

21
When monitoring a patient, one Philips M4607A rechargeable Lithium Ion battery must always be
inserted into the battery compartment on the right side of the X2. This applies even when you are
running the X2 external power, either via the external power supply or when connected to a host
monitor. The battery has the effect of sealing the battery compartment, thereby preventing the ingress
of fluids or foreign bodies. A severe yellow INOP (!!INSERT BATTERY) will be issued if the
monitor is connected to AC mains without a battery fully inserted in the battery compartment. This
INOP will persist until a battery is loaded.
Battery compartment
To use the X2 with battery power, disconnect it from the host monitor or the external power supply
(M8023A).
You can switch between battery-powered and mains-powered (AC) operation without interrupting
monitoring.
When an X2 is connected to a host monitor, no battery data from the X2 is visible on the host. Battery
INOPs, however, are displayed.
The battery recharges automatically when the X2 is connected to a host monitor (running either on
AC mains or battery power) or the optional external power supply (M8023A).
221