Patient Alarms And Inops; Patient Alarm Messages - Philips IntelliVue X2 Instructions For Use Manual

Multi-measurement module
Hide thumbs Also See for IntelliVue X2:
Table of Contents

Advertisement

Patient Alarms and INOPs

5
This chapter lists patient alarms and technical alarms (INOPs) alphabetically, irrespective of their
priority. INOPs start on page 61. All alarms and INOPs are listed here; the ones which can appear on
your monitor will depend on the model and the individual options.

Patient Alarm Messages

The measurement labels and abbreviations for pressure, temperature, SpO
alarms are explained in the individual chapters.
Some alarms may be shown at the Information Center in shortened form, when transferred through
IntelliVue Instrument Telemetry. These shortened alarm texts are included in the list and identified
with the note "at Information Center".
Note that yellow arrhythmia alarms may be shown with one or with two stars, depending on your
monitor configuration and the Information Center revision you are using.
Refer to your IntelliBridge Device Driver Instructions for Use for patient alarms from connected
external devices.
Alarm Message
*/**AFIB
***APNEA or
***APNEA mm:ss sec
or
***APNEA >10min
***ASYSTOLE
**awRR HIGH
**awRR LOW
**BIS HIGH
From
Condition
ECG/
Atrial fibrillation waveform detected
Arrhythmia
CO
, Resp,
Respiration has stopped for longer than
2
Spirometry
the preset apnea time. "mm:ss" denotes
the Apnea duration in minutes and
seconds.
ECG
No QRS detected for a period greater
than the asystole threshold (in the
absence of Vfib or chaotic ECG).
CO
, Resp,
The airway respiration rate has
2
AGM
exceeded the high alarm limit.
CO
, Resp,
The airway respiration rate has dropped
2
AGM
below the low alarm limit.
BIS
The Bispectral Index value has
exceeded the high alarm limit.
5
, and anesthetic agent
2
Indication
yellow alarm lamp, short
yellow audible alarm.
numeric flashes, red alarm
lamp, alarm tone.
numeric flashes, red alarm
lamp, alarm tone.
numeric flashes and high limit
is highlighted, yellow alarm
lamp, alarm tone.
numeric flashes and low limit
is highlighted, yellow alarm
lamp, alarm tone.
numeric flashes and high limit
is highlighted, yellow alarm
lamp, alarm tone.
55

Advertisement

Table of Contents
loading

Table of Contents