Philips IntelliVue X2 Instructions For Use Manual page 68

Multi-measurement module
Hide thumbs Also See for IntelliVue X2:
Table of Contents

Advertisement

5 Patient Alarms and INOPs
Alarm Message
*/**/***TELE ALARM
**<Temperature
label> HIGH
**<Temperature
label> LOW
*/**VENT BIGEMINY
***VENT FIB/TACH
*/**VENT RHYTHM
*/**VENT TRIGEMINY
***VTACH
**/***VueLink ALARM
at Information Center
60
From
Condition
Telemetry
This is a generic alarm from the
telemetry system. The specific alarm
cause is indicated in the alarm message
in the Telemetry Data Window.
TEMP
The temperature has exceeded the high
alarm limit.
TEMP
The temperature has fallen below the
low alarm limit.
ECG/
A dominant rhythm of N, V, N, V (N =
Arrhythmia
supraventricular beat, V = ventricular
beat) was detected.
ECG
A fibrillatory waveform for 4
consecutive seconds was detected.
ECG/
A dominant rhythm of adjacent Vs >
Arrhythmia
vent rhythm limit and ventricular HR <
VTach HR limit was detected.
ECG/
A dominant rhythm of N, N, V, N, N,
Arrhythmia
V (N = supraventricular beat, V =
ventricular beat) was detected.
ECG,
Ventricular tachycardia has been
Arrhythmia
detected (Consecutive PVCs exceed V-
Tach Run limit and HR exceeds V-Tach
HR limit).
VueLink
A yellow (**) or red (***) patient alarm
is present on the VueLink module.
Check the monitor display for more
detailed alarm information.
Patient Alarm Messages
Indication
yellow or red alarm lamp and
alarm tone
numeric flashes and high limit
is highlighted, yellow alarm
lamp, alarm tone.
numeric flashes and low limit
is highlighted, yellow alarm
lamp, alarm tone.
numeric flashes, yellow alarm
lamp, short yellow audible
alarm.
numeric flashes, red alarm
lamp, alarm tone.
numeric flashes, yellow alarm
lamp, short yellow audible
alarm.
numeric flashes, yellow alarm
lamp, short yellow audible
alarm.
numeric flashes, red alarm
lamp, alarm tone.
(on monitor) yellow or red
alarm lamp, alarm tone