Technical Alarm Messages (INOPs)
INOP Message, Indication
Check Screen Res
INOP tone
Check Settings
INOP tone
Check SpeedPoint
INOP tone.
Check Touch Input
INOP tone
Check Waves
INOP tone
CHK ECG Sync Cable
INOP tone
Chk IndepDsp Cable
CHK MSL Connection
INOP tone
Chk SpO
2
INOP tone
C LEAD OFF
HR Numeric is replaced by -?- for 10
seconds. INOP tone.
CO
AUTO ZERO
2
Numeric is replaced by a -?-
if the Autozero lasts >15 sec, INOP
tone sounds.
CO
CAL MODE
2
CO
numeric displays current CO
2
value for accuracy check
CO
CAL RUNNING
2
Numeric is replaced by a -?-
CO
CHANGE SCALE
2
CO
CHK ADAPTER
2
Numeric is replaced by a -?-
INOP tone.
CO
CHECK CAL
2
Numeric is replaced by a -?-
INOP tone.
Source
Monitor
Monitor
Monitor
Monitor
Monitor
Monitor
Monitor
Monitor
Telemetry
T Settings
ECG
CO
CO
2
CO
CO
CO
CO
What to do
The Screen you have selected uses a resolution which is not
supported by the display. The monitor will show a generic
Screen instead until you select a different Screen.
Contact your service personnel if you want the Screen deleted
from the Profile(s) to avoid this in future.
If this INOP appears, check the monitor and patient settings
before you resume monitoring. If the settings are unexpected,
there may be a problem with the monitor software. Contact
your service personnel.
Perform a visual and functional check of the SpeedPoint input
device. Contact your service personnel.
Perform a visual and functional check of the touch input
device. Contact your service personnel.
The options purchased with this monitor may not support the
number of waves required to show the selected Screen, so some
waves or high resolution trends are missing from the Screen.
Select a different Screen with fewer waves.
Contact your service personnel if you want the Screen deleted
from the Profile(s) to avoid this in future.
The ECG Sync is detecting an invalid signal, or the ECG Sync
cable is disconnected.
The monitor cannot communicate with the D80 Intelligent
Display. Check the MSL coupling cable. The end with the grey
connector must be connected to the Intelligent Display.
Check that the MSL connector or cable are properly
connected. Check the cable and connector for damage.
Synchronization of SpO
Information Center has failed. Check that the SpO
in use are appropriate.
The C electrode (AAMI: V electrode) has become detached
from the patient or the lead set has been changed. Reattach the
electrode or select
New Lead Setup
menu to confirm the new lead set.
The automatic zero calibration is in progress. This typically
2
takes 10 seconds. During this time the CO
updated, or they may be replaced by -?-. Wait until the zero
calibration is complete to resume monitoring.
Currently no calibration is running. Accuracy can be checked
2
by placing the transducer on the two cells of the calstick and
starting calibration. To start monitoring, leave Cal. Mode.
Wait until calibration is finished.
2
The CO
wave is clipped. Select a more appropriate wave scale
2
2
to display the whole wave.
Check that the sensor is connected to the airway adapter, clean
2
the airway adapter, if necessary. Perform a zero calibration. If
the INOP persists, contact your service personnel.
The CO
value is outside the measurement range. Perform an
2
2
accuracy check for both calstick cells and, if necessary,
recalibrate the transducer.
5 Patient Alarms and INOPs
T settings between the monitor and
2
T settings
2
in the
Setup ECG
values may not be
2
67