Extending Measurements
M3014A, M3015A and M3016A Measurement Extensions
When attached to an X2 connected to external power (either via the external power supply or a host
monitor), the optional M3014A Capnography extension adds mainstream capnography, and
optionally one pressure plus either a pressure or a temperature to the monitor. When the X2 is
connected to a host monitor, Cardiac Output (C.O.) measurements can be performed and can be used
to calibrate Continuous Cardiac Output (CCO). These stored C.O. results and the calibrated CCO
measurement are still available in the X2 when it is disconnected from the host monitor.
The optional M3015A Microstream CO
either pressure or temperature to the monitor. The optional M3016A Mainstream CO
mainstream capnography and optionally either pressure or temperature to the monitor.
Only one CO
4
3
1
Pressure connectors (red)
2
Temperature connector (brown)
3
Mainstream/sidestream connector CO
(optional)
4
Cardiac Output connector
measurement at a time is supported.
2
M3014A Capnography
3
M3016A Mainstream
extension adds microstream capnography and optionally
2
1
7
2
1
2
5
Inlet
Microstream
6
7
Gas sample outlet
2
2 Basic Operation
extension adds
2
M3015A Microstream
1
5
6
connector CO
2
2
15