Monitor Performance Specifications
X2 (M3002A) Performance Specifications
Alarm signal
Review Alarms
Real Time Clock
Buffered Memory
M8023A External Power Supply Performance Specifications
Power Specifications
Indicators
X2 (M3002A) Interface Specifications
Measurement Link
(MSL)
ECG Sync Pulse
Output
(See "Connectors" on
page 249 for safety-
related information)
System delay
less than 3 seconds
Pause duration
1,2,3 minutes or infinite, depending on configuration
Extended alarm pause
5 or 10 minutes
Information
all alarms / inops, main alarms on/off, alarm silence and time
of occurrence
Capacity
300 items
Range
from: January 1, 1997, 00:00 to: December 31, 2080, 23:59
Accuracy
<4 seconds per day (typically)
Hold Time
infinite if powered by host monitor or external power supply;
otherwise at least 48 hours
Hold Time
if powered by host monitor or external power supply: infinite
without power: at least 48 hours (typical: >72 hours)
Contents
Active settings, trends, patient data, realtime reports, review
alarms
Power consumption
<12 W average
<30 W peak
Line Voltage
100 to 240 V ~
Current
1.3 to 0.7 A
Frequency
50/60 Hz ~
AC Power
green LED
Connectors
Female ODU (Proprietary)
Power
30 V to 60 V input
Power Sync.
RS-422 compliant input 78.125 kHz (typical)
LAN signals
IEEE 802.3 10-Base-T compliant
Serial signals
RS-422 compliant
Local signals
Provided for connecting MMS extensions
Cable detection
Yes
Marker In
No
Wave Output
No
Connector
Binder Series 709/719
Output levels
Output low <0.8 V @ I = -4 mA
Output high >2.4 V @ I = 4 mA
Isolation
None
Pulse Width
100 +/- 10 ms (high)
Delay from R-wave
20 ms maximum per AAMI EC13
peak to start of pulse
Minimum required R-
0.5 V
wave amplitude
24 Specifications
255