Philips IntelliVue X2 Instructions For Use Manual page 109

Multi-measurement module
Hide thumbs Also See for IntelliVue X2:
Table of Contents

Advertisement

ECG Lead Placements
Standard 3-Lead Placement
Standard 5-Lead Placement
To Purchase, Visit
Avobus.com
Electrode labels
AAMI
EASI
IEC
V3
C3
V4
C4
V5
C5
V6
C6
1
3
1
V
5
3
or call
1-800-674-3655
7 ECG, Arrhythmia, ST and QT Monitoring
Electrode colors
AAMI
IEC
Brown/Green
White/Green
Brown/Blue
White/Brown
Brown/Orange
White/Black
Brown/Violet
White/Violet
1
RA placement: directly below the clavicle
and near the right shoulder
2
LA placement: directly below the
2
clavicle and near the left shoulder
3
LL placement: on the left lower
abdomen
RA placement:
near the right shoulder
LA placement:
2
near the left shoulder
RL placement:
LL placement:
V placement:
on your required lead selection
4
directly below the clavicle and
directly below the clavicle and
on the right lower abdomen
on the left lower abdomen
on the chest, the position depends
101