Philips IntelliVue X2 Instructions For Use Manual page 66

Multi-measurement module
Hide thumbs Also See for IntelliVue X2:
Table of Contents

Advertisement

5 Patient Alarms and INOPs
Alarm Message
***<Pressure>
DISCONNECT
***<Pressure> HIGH
**<Pressure> HIGH
***<Pressure> LOW
**<Pressure> LOW
**Pulse HIGH
**Pulse LOW
*/**PVCs/min HIGH
**QTc HIGH
**ΔQTc HIGH
*/**R-ON-T PVCs
**RR HIGH
58
From
Condition
PRESS
The pressure is non-pulsatile and the
mean pressure is continuously less than
10mmHg (1.3kPa). This alarm occurs
only with arterial pressures (P, ABP,
ART, Ao, BAP, FAP, PAP, UAP, P1, P2,
P3, P4).
PRESS
The measured pressure value is above
the extreme high alarm limit. s, d, or m
after the label indicates whether the
systolic, diastolic or mean pressure has
crossed the limit.
PRESS
The measured pressure value is above
the high alarm limit. s, d, or m after the
label indicates whether the systolic,
diastolic or mean pressure has crossed
the limit.
PRESS
The measured pressure value is below
the extreme low alarm limit. s, d, or m
after the label indicates whether the
systolic, diastolic or mean pressure has
crossed the limit.
PRESS
The measured pressure value is below
the low alarm limit. s, d, or m after the
label indicates whether the systolic,
diastolic or mean pressure has crossed
the limit.
PRESS
The pulse rate has exceeded the high
SpO
alarm limit.
2
PRESS
The pulse rate has dropped below the
SpO
low alarm limit.
2
ECG/
More premature ventricular
Arrhythmia
contractions have been detected in a
minute than the limit.
ECG/QT
QTc value has exceeded the QTc high
limit for more than 5 minutes
ΔQTc value has exceeded the ΔQTc
ECG/QT
high limit for more than 5 minutes
ECG/
For HR <100, a PVC with R-R interval
Arrhythmia
< 1/3 the average interval followed by a
compensatory pause of 1.25 x average
R-R interval or two such Vs without
compensatory pause occurring within 5
minutes of each other. (When HR
>100, 1/3 R-R interval is too short for
detection.).
RESP
The respiration rate has exceeded the
high alarm limit.
Patient Alarm Messages
Indication
numeric flashes, red alarm
lamp, alarm tone.
numeric flashes, high limit is
highlighted, red alarm lamp,
alarm tone.
numeric flashes, high limit is
highlighted, yellow alarm
lamp, alarm tone.
numeric flashes and low limit
is highlighted, red alarm lamp,
alarm tone.
numeric flashes and low limit
is highlighted, yellow alarm
lamp, alarm tone.
numeric flashes and high limit
is highlighted, yellow alarm
lamp, alarm tone.
numeric flashes and low limit
is highlighted, yellow alarm
lamp, alarm tone.
numeric flashes, yellow alarm
lamp, short yellow audible
alarm.
numeric flashes, yellow alarm
lamp, alarm tone.
numeric flashes, yellow alarm
lamp, alarm tone.
numeric flashes, yellow alarm
lamp, short yellow audible
alarm.
numeric flashes and high limit
is highlighted, yellow alarm
lamp, alarm tone.