5 Patient Alarms and INOPs
INOP Message, Indication
<Temp label> OVERRANGE
Numeric is replaced by -?-
INOP tone.
Tesoph INOPS
TEXT UPLOAD FAILED
INOP tone
TimeExpired:<timer label>
INOP tone
Tnaso INOPS
Trect INOPS
Tskin INOPS
Ttymp INOPS
Tven INOPS
Tvesic INOPS
UAP INOPS
Unsupported LAN
INOP tone
User I/F Malfunct.
INOP tone.
UVP INOPS
V LEAD OFF
Numeric is replaced by -?- for 10
seconds; INOP tone.
<VueLink option> CHK
CABLE
INOP tone.
<VueLink option> CHK
CONF.
INOP tone.
82
Source
What to do
TEMP
Try changing the application site of the transducer.
[The temperature is less than -1
TEMP
See <Temp label> INOPs (under Temp).
IntelliBridge Incoming text from the IntelliBridge modules exceeds the
maximum limit. Try unplugging one of the IntelliBridge
modules. If the INOP occurs repeatedly contact your service
personnel; a software upgrade may be necessary.
Monitor
The time has expired for the timer indicated in the INOP text.
Clearing the timer clears the INOP.
TEMP
See <Temp label> INOPs (under Temp).
TEMP
See <Temp label> INOPs (under Temp).
TEMP
See <Temp label> INOPs (under Temp).
TEMP
See <Temp label> INOPs (under Temp)
TEMP
See <Temp label> INOPs (under Temp).
TEMP
See <Temp label> INOPs (under Temp)
PRESS
See <Pressure label> INOPS (under Pressure).
Monitor
There is a problem with the communication to the network
and central monitoring is currently not possible. Check the
connection. If the INOP persists, switch off the monitor and
contact your service personnel.
Monitor
Perform a visual and functional check of all the monitor input
devices. Contact your service personnel.
PRESS
See <Pressure label> INOPS (under Pressure).
ECG
The V electrode (IEC: C electrode) has become detached from
the patient or the lead set has been changed. Reattach the
electrode or select
menu to confirm the new lead set.
VueLink
No cable or the wrong cable connected to the VueLink module,
or incorrect device selected. Silencing this INOP switches the
measurement off.
VueLink INOP abbreviations may differ slightly depending on
the device category.
VueLink
The wrong external device has been selected on the VueLink
module, or the external device has not been correctly setup, or
the wrong cable has been used to connect the device to the
VueLink module.
VueLink INOP abbreviations may differ slightly depending on
the device category.
Technical Alarm Messages (INOPs)
°
C, or greater than 45
New Lead Setup
in the
Setup ECG
°
C.]