Ecg Lead Fallback; Ecg Lead Placements - Philips IntelliVue X2 Instructions For Use Manual

Multi-measurement module
Hide thumbs Also See for IntelliVue X2:
Table of Contents

Advertisement

7 ECG, Arrhythmia, ST and QT Monitoring
ECG Leads Monitored
If you are using
a 3-electrode set
a 5-electrode set
a 6-electrode set
a 10-electrode set
an EASI 5-electrode set I, II, III, aVR, aVL, aVF, V1, V2, V3, V4, V5, V6 I and A
Changing Lead Sets
To change the ECG lead set,
• When adding leads: place the additional electrodes as required - the monitor automatically
recognizes the new lead placement.
• When removing leads: there are two possibilities - removing the complete chest leads block or
removing individual leads. If you unplug the chest lead block from the trunk cable connector, the
monitor will automatically recognize the new lead placement. Remove individual leads by removing
the electrodes. The monitor will issue a Lead Off INOP message; select New Lead Setup in
the Setup ECG menu and the INOP message will disappear.

ECG Lead Fallback

If fallback is configured on and there is a leads off INOP in the primary lead (and in the secondary
lead, if you are using multi-lead monitoring) for longer than 10 seconds, and if another lead is
available, this available lead automatically becomes the primary lead. This is known as lead fallback.
When the Leads Off condition is corrected, the leads are automatically switched back.
This setting can only be changed in Configuration Mode.

ECG Lead Placements

The labels and colors of the ECG electrodes differ according to the standards that apply for your
hospital. The electrode placement illustrations in this chapter use the AAMI labels and colors.
Electrode labels
AAMI
RA
LA
LL
RL
V
V1
V2
100
these leads are available:
I, II, III
I, II, III, aVR, aVL, aVF, V and MCL
I, II, III, aVR, aVL, aVF, Va, Vb
I, II, III, aVR, aVL, aVF, V1, V2, V3, V4, V5, V6 RA and LL
Electrode colors
EASI
IEC
AAMI
I
R
White
S
L
Black
A
F
Red
N
N
Green
E
C
Brown
C1
Brown/Red
C2
Brown/Yellow
ECG Lead Fallback
Resp is measured
between electrodes:
RA and LL
RA and LL
RA and LL
IEC
Red
Yellow
Green
Black
White
White/Red
White/Yellow