Philips IntelliVue X2 Instructions For Use Manual page 184

Multi-measurement module
Hide thumbs Also See for IntelliVue X2:
Table of Contents

Advertisement

14 Monitoring Carbon Dioxide
3
4
5
6
Attaching and Removing the CO
1
2
WARNING
To prevent stress on the endotrachial tube, support the transducer and airway adapter.
To avoid infection, use only sterilized airway adapters.
176
To Purchase, Visit
Avobus.com
Enter the calibration value printed on the calstick then press
When the message CO
CAL1 done;Start CAL2 cal appears, put the transducer on the
2
other cell and select Start CAL2 then press
When you see the message CO
Select Cal. Mode to switch calibration mode off. You cannot monitor in calibration mode.
Open the latch and place the
transducer onto the airway adapter.
Place the airway adapter in the
patient's breathing circuit between
the endotracheal tube and the Y-
piece. You may see the CO
2
SENSOR WARMUP message until
the transducer reaches operating
temperature. Wait until this
disappears before starting the
measurement.
To remove the transducer from the
airway adapter, open the latch and
pull out the airway adapter.
or call
1-800-674-3655
Measuring Mainstream CO
.
Confirm
cal done;leave Cal. Mode, calibration is complete.
2
Transducer
2
Airway Adapter
CO
Transducer
2
using M3016A
2
to start calibration.
Confirm