15 Assigning Two Devices to One Patient
Correct channel configuration is important, refer to the Configuration Guide for details.
If a TELE DISCONNECT INOP occurs when the telemetry device goes out-of-range, the INOP will
disappear (without being silenced) as soon as the ECG signal from the telemetry device is available at
the Information Center via standard telemetry transmission.
SpO
T Manual Mode
2
When a telemetry device with a short range radio connection supplies SpO
monitor will have a manual measurement mode available. This allows SpO
on request and not continuously, helping to save battery power. To ensure there is no gap in SpO
measurements when moving from standard telemetry transmission to short range radio transmission,
the measurement mode will always be set to continuous in this situation unless manual mode is set in
both the telemetry device and the monitor.
Unassigning a Telemetry Transceiver with SRR Adapter at the Monitor
If a monitor and a telemetry device are paired, the assignment and the short range radio connection
will be ended when the devices are unpaired (see "Unpairing the Monitor and Telemetry Device" on
page 184).
If the monitor does not have a connection to an Information Center, you must manually end the
assignment and the short range radio connection.
Select the Measurement Selection key.
1
In the Measurement Selection window, select the Tele pop-up key.
2
In the Tele device window select Unassign Tele.
3
Pairing an X2 With a Telemetry Interface to a Host Monitor
An X2 using an IntelliVue Instrument Telemetry interface will be automatically paired when it is
connected to a host monitor and it has been declared as a "telemetry device" at the Information Center.
When the X2 is disconnected from the host monitor it will remain paired and the measurement data
will appear at the host monitor in the Telemetry Data window. After the X2 is disconnected from the
monitor it may take over 15 seconds until its data is displayed at the Information Center.
N O T E
Pairing of a host monitor and a telemetry device, with all the related functionality, is only possible
when the host monitor is using a wired LAN connection or a wireless LAN interface (but not using the
IntelliVue Instrument Telemetry interface).
The X2 can also be paired to a host monitor without a direct connection, as described in "Indirect
Connection - Manual Pairing" on page 181.
Refer also to "Use Models With Telemetry" on page 188 for further related use modes.
Unpairing the Monitor and Telemetry Device
If the patient will no longer be monitored with the telemetry device, or only with the telemetry device
and no longer with the monitor, you need to end the device pairing. After unpairing, the Information
Center will receive data exclusively from the monitor or from the telemetry device:
select the Unpair To Mon. SmartKey to end pairing and have the Information
Center receive the measurement data from the monitor
184