Technical Alarm Messages (INOPs)
INOP Message, Indication
<pTemp label> CHECK PROBE
Numeric is replaced by -?-
INOP tone.
<pTemp label> DEACTIVATED
INOP tone.
<pTemp label> EQUIP MALF
Numeric is replaced by -?-
INOP tone.
<pTemp label> INCOMPAT.
<pTemp label> MEAS FAILED
Numeric is replaced by -?-
INOP tone.
<pTemp label> NO PROBE
Numeric is replaced by -?-
INOP tone.
<pTemp label> OVERRANGE
Numeric is replaced by -?-
INOP tone.
<pTemp label> PROBE MALF
Numeric is replaced by -?-
INOP tone.
pTaxil INOPS
pToral INOPS
pTrect INOPS
(!!)PW:Action Required
PW:Check Settings
PW in conflict
RA LEAD OFF
Numeric is replaced by -?-
INOP tone.
RAP INOPS
Rem.Alarmdev.Malf
INOP tone
Rem. Display Malf.
Source
What to do
Predictive
Check that the probe holder is correctly installed and that a
Temp
compatible probe is in use. If the INOP does not clear, remove
the probe from the holder then replace it. If the INOP still does
not clear, disconnect and reconnect the probe.
Predictive
The Predictive temperature has been deactivated.
Temp
Predictive
The Predictive Temperature hardware is defective. Contact
Temp
your service personnel.
Predictive
The Predictive Temperature version is not supported by the
Temp
monitor.
Predictive
The ambient temperature is outside the specified range (10 to
Temp
40°C /50 to 104°F). Bring the temperature into range to
continue monitoring. If the INOP does not clear, disconnect
and reconnect the probe.
Predictive
Connect a probe to the Predictive temperature unit.
Temp
Predictive
The temperature at the measurement site is out of range. Check
Temp
that the probe is correctly located. If the INOP does not clear,
disconnect and reconnect the probe.
Predictive
The connected probe may be defective. Disconnect and
Temp
reconnect the probe. If the INOP does not clear, try another
probe.
Predictive
See <pTemp label> INOPS
Temp
Predictive
See <pTemp label> INOPS
Temp
Predictive
See <pTemp label> INOPS
Temp
Protocol
The protocol currently running requires a user response. Check
Watch
which pop-up window is displayed and provide the appropriate
response.
Protocol
Contact your service personnel. Settings could not be loaded or
Watch
interpreted correctly
Protocol
There is a patient information mismatch which has not yet
Watch
been resolved (>15 minutes).
ECG
The RA electrode has become detached from the patient or the
lead set has been changed. Reattach the electrode or select
Lead Setup
new lead set.
PRESS
See <Pressure label> INOPS (under Pressure).
Monitor
There is a problem with the connection to the remote alert
device. Contact your service personnel to check the remote
alert device and its connections.
Remote
There is a problem with an input device at the remote display.
display
Perform a visual and functional check of all input devices.
Contact your service personnel.
5 Patient Alarms and INOPs