24 Specifications

Measurement Specifications

See the Appendix on Default Settings for a list of the settings the monitor is initially shipped with.
ECG/Arrhythmia/ST/QT
Complies with IEC 60601-2-25:1993 + A1:1999 /EN60601-2-25:1995 + A1:1999, IEC 60601-2-
27:2005/EN60601-2-27:2006, IEC 60601-2-51:2003 /EN 60601-2-51:2003 and AAMI EC11/
EC13:1991/2002.
ECG/Arrhythmia/ST Performance Specifications
Cardiotach
PVC Rate
ST Numeric
QT Numeric
QTc Numeric
ΔQTc Numeric
QT-HR Numeric
Sinus and SV Rhythm
Ranges
Bandwidth
258
Range
Adult/pedi: 15 to 300 bpm
Neo range: 15 to 350 bpm
Accuracy
±1% of range
Resolution
1 bpm
≥200 µV
Sensitivity
peak
Range
0 to 300 bpm
Resolution
1 bpm
Range
-20 to +20 mm
Accuracy
±0.5 mm or 15%, whichever is greater
Resolution
0.1 mm
Range
200 to 800 ms
Accuracy
±30 ms
Resolution
8 ms
Range
200 to 800 ms
Resolution
1 ms
Range
-600 to +600 ms
Resolution
1 ms
Range - adult
15 to 300 bpm
Range - pediatric
15 to 350 bpm
and neonatal
Brady
Adult: 15 to 59 bpm
Pedi: 15 to 79 bpm
Neo: 15 to 89 bpm
Normal
Adult: 60 to 100 bpm
Pedi: 80 to 160 bpm
Neo: 90 to 180 bpm
Tachy
Adult: >100 bpm
Pedi: >160 bpm
Neo: >180 bpm
Diagnostic Mode
Adult/neo/pedi: 0.05 to 150 Hz
Extended
Neo/pedi: 0.5 to 150 Hz
Monitoring Mode
Monitoring Mode Adult: 0.5 to 40 Hz
Neo/pedi: 0.5 to 55 Hz
Filter Mode
Adult/neo/pedi: 0.5 to 20 Hz