Technical Alarm Messages (INOPs)
INOP Message, Indication
<SpO
LABEL> PULSE?
2
Numeric is replaced by -?-
INOP tone
<SpO
LABEL> SEARCHING
2
Numeric unavailable
<SpO
label> SENSOR MALF
2
Numeric is replaced by -?-
INOP tone.
<SpO
LABEL> SENSOR OFF
2
Numeric is replaced by -?-
INOP tone
<SpO
LABEL> UNKN.SENSOR
2
Numeric is replaced by a -?-
<SpO
LABEL> UNPLUGGED
2
Numeric is replaced by -?-
INOP tone
<SpO
LABEL> UPGRADE
2
Label is displayed with a -?-, numeric is
unavailable
Sp - vO
2
Numeric is replaced by -?-
Sp - vO
2
Numeric is replaced by -?-
SRR INTERFERENCE
INOP tone
SRR INVALID CHAN
INOP tone
SRR MALFUNCTION
SVR/SVRI CHK SOURCES
Numeric is replaced by -?-
SVR/SVRI CHK UNITS
Numeric is replaced by -?-
SVR/SVRI SET CVP USED
Numeric is replaced by -?-
T/T1/T2/T3/T4 INOPs
TAAP DISABLED
INOP tone
Tamb INOPs
Tart INOPs
Source
SpO
SpO
SpO
SpO
SpO
SpO
SpO
CHK SOURCES
Sp - vO
CHK UNITS
Sp - vO
Monitor
Monitor
SVR/SVRI
SVR/SVRI
SVR/SVRI
TEMP
Monitor
TEMP
TEMP
What to do
The detectable pulsations of the SpO
2
specified pulse rate range.
SpO
is analyzing the patient signal to derive Pulse, SpO
2
2
Perf values. Please wait until the search analysis is complete.
The SpO
sensor or adapter cable is faulty. Try another adapter
2
2
cable and sensor. If the INOP persists, contact your service
personnel.
The SpO
sensor is not properly applied to the patient. Apply
2
2
the sensor following the instructions supplied by the
manufacturer.
The connected sensor or adapter cable is not supported by the
2
SpO
measurement. Use only specified sensors and cables.
2
An SpO
measurement label has been deactivated, either by
2
2
unplugging a module, or by deactivating the label in the
Measurement Selection window.
The measurement automatically disappears from the display.
To switch the measurement on again, either replug the module
or reactivate the measurement label in the Measurement
Selection window.
The SpO
measurement is currently in UPGRADE mode.
2
2
Monitoring is not possible in this mode.
Not all measurements or values required to perform the
2
calculation are available. Check measurement sources.
The monitor has detected a conflict in the units used for this
2
calculation. Check the unit settings.
The short range radio connection has interference from
another device. Try using another channel.
The channel configuration of the Short Range Radio is invalid.
Check channel and channel mask configuration.
Malfunction in the short range radio device. If the INOP
persists contact your service personnel.
Not all measurements or values required to perform the
calculation are available. Check measurement sources.
The monitor has detected a conflict in the units used for this
calculation. Check the unit settings.
A CVP value is required for this calculation, but is not
currently being measured. The monitor is using the CVP value
Setup SVR
preset in the
See <Temp label> INOPs (under Temp)
The currently selected telemetry configuration on the monitor
does not allow connection of telemetry devices to the monitor.
See <Temp label> INOPs (under Temp)
See <Temp label> INOPs (under Temp)
5 Patient Alarms and INOPs
signal are outside the
2
menu.
and
2
79