Cylinder Information
• Only use a size "88" (1,963 liters) cylinder that is marked
WARNING:
"MR Conditional. Keep cylinder at 100 gauss or less." with the
INOmax DS
other cylinder may create a projectile hazard.
• Only use manufacturer supplied drug cylinders, regulators and
adapters.
• Cylinders should be stored between 59-86 degrees F
(15-30 degrees C).
• Always secure a cylinder when not using it.
• Never lift a cylinder by its valve.
• Never drop a cylinder.
• Never use a hammer, pry or wedge to loosen a valve or
protection cap. The valve and protection cap should be operated
by hand.
• Never let oil, grease or other combustibles come in contact with
a cylinder or valve.
• Never remove or deface cylinder labeling or markings.
• Never attempt to repair a leaking cylinder valve or its safety relief device.
• Never operate equipment that is leaking.
• Never ship a leaking cylinder.
• Never store cylinders:
- where damage can result from the elements, such as standing water or
temperatures over 125 degrees F (52 degrees C).
- where they can contact corrosive substances.
- where they can be cut or abraded by an object.
- next to a walkway, elevator or platform edge.
Purging the INOmax DS
• All INOmax DS
patient does not receive an excess level of NO
• If the INOmax DS
minutes, depressurize the regulator supply line.
• If the INOmax DS
minutes, repeat automated or manual purge procedure.
• If the INOmax DS
within 12 hours, repeat pre-use procedure.
• Use a properly designed cart to move a cylinder and properly secure the cylinder when moving it.
Note:
• Apply a proper pressure regulating device to the cylinder before using it.
• Periodically check the cylinder pressure.
• Apply the valve outlet cap and valve protective cap to a cylinder when it is not connected.
Part No. 20568 Rev-02
2015-08
Plus MRI while in the scanner room. Use of any
IR
Plus MRI
IR
Plus MRI devices must be purged before use to ensure the
IR
Plus MRI is not going to be used on a patient within 10
IR
Plus MRI is not used and is pressurized for more than 10
IR
Plus MRI is depressurized and not used
IR
Size 88 Size D
.
2
3-9