Oxygen Sensor High Calibration - Mallinckrodt INOmax DSIR Plus MRI Operation Manual

Table of Contents

Advertisement

Oxygen Sensor High Calibration

Do not perform a high calibration procedure in the MR scanner room. Calibration
WARNING:
equipment is a potential projectile hazard.
Never connect the sample line directly to a high pressure gas source (greater than
Caution:
150 cmH
2
Note:
Complete a low calibration (see Low Calibration section) prior to completing the high calibration.
The oxygen high calibration requires a user supplied source of 100% oxygen.
1
2
3 4
3c
3b
5-4
O); this could damage the sampling system.
5
3a
1. From the high calibration screen
(third menu level), press the
100% O
button to initiate the O
2
high calibration wizard.
If the date is flashing, it signifies the
Note:
calibration is past due.
2. Assemble connectors into a calibration setup
and set oxygen flow to five L/min.
Source
(1) 100% O
2
(2) O
Tubing
2
(3) 15M X 4.5 mm I.D Adapter
(4) Gas Sample Tee
(5) Patient Gas Sample Line with Nafion
3a. To continue the high calibration wizard
press the NEXT button.
3b. To start the high calibration without using
the wizard, press the START CAL button
3c. To exit the O
high calibration wizard, press
2
the CANCEL button.
2
Part No. 20568 Rev-02
2015-08