Alarm History

Recent Alarms button
Figure 4-3 Recent Alarms Button on the Main Screen
Clear Recent Alarms button
Figure 4-4 Recent Alarms Screen
Part No. 20568 Rev-02
2015-08
When an alarm condition has been resolved, the
alarm message is no longer displayed on the main
screen.
The recent alarms can be seen by
pressing the Recent Alarms button.
The Recent Alarms button is present and displayed
as a "double-bell" when there are no active alarms
and any previously resolved alarms have not been
cleared.
The alarms are displayed in chronological order,
with most recent at the top (for example, Figure 4-4
the recent alarm conditions that have occurred).
• Alarm history dates and times are
Note:
displayed per user-set off-set time
(see Time Adjust setting, page 1-18).
• Alarm conditions lasting less than one
second may not display in the alarm
area of the user screen, but will post to
the alarm log and alarm history.
Press the clear Recent Alarms
button to clear the recent alarms
and return to the main screen.
To return to the main screen
without clearing the recent alarm
history, press the EXIT or the
Return to Previous Menu button.
If no action is taken, the system
will return to the main screen after
30 seconds.